A Uterine Suction Device to Prevent Postpartum Hemorrhage in Scheduled Cesarean Delivery: A Pilot Study

Postpartum hemorrhage (PPH) remains a leading cause of maternal morbidity and mortality, with cesarean delivery posing a heightened risk. While interventions such as prophylactic tranexamic acid and balloon tamponade have limitations-especially when the cervix is not dilated-vacuum-assisted uterine tamponade may offer a novel intraoperative approach. This prospective pilot study evaluated the feasibility of the Daisy catheter, a cervical drain device developed by Raydiant Oximetry, Inc., designed to evacuate blood and promote uterine contraction through continuous negative-pressure suction.We enrolled ten pregnant individuals scheduled for cesarean delivery at a tertiary care center, all of whom had at least one PPH risk factor. Following fetal and placental delivery, the Daisy catheter was inserted trans-hysterotomically, advanced through the cervix, and connected to wall suction (-90 to -100 mm Hg) for 2 hours. Quantitative blood loss, perioperative hemoglobin change, ultrasound findings, and adverse events were recorded.Device placement succeeded in 9 of 10 cases; one failure was due to an undiagnosed cervical cerclage. Eight participants completed the full suction protocol. Mean hemoglobin decline from preoperative baseline to postoperative day 1 was 1.36 ± 0.47 g/dL, significantly lower than the 1.9 ± 1.1 g/dL observed in a historical cohort from 31 U.S. hospitals (<i>p</i> = 0.019). Ultrasound at 2 hours postpartum confirmed correct device placement, absence of intrauterine clot, and no evidence of trauma. Device removal was uncomplicated, and no adverse events were reported.These preliminary findings suggest that intraoperative use of the Daisy device is feasible, well-tolerated, and may reduce blood loss after cesarean delivery. Larger, randomized trials are warranted to evaluate its impact on transfusion rates, reoperation, and overall maternal outcomes, particularly in settings where alternative tamponade methods are limited. The ClinicalTrials.gov identifier is NCT06219538. · Vacuum tamponade used during cesarean delivery.. · Daisy device showed safe, feasible deployment.. · Hemoglobin drop was lower than the historical average..

Authors

Institutions

Publication Details

Journal
American Journal of Perinatology
Published
2025-11-24
DOI
https://doi.org/10.1055/a-2752-8798
Primary Topic
Maternal and fetal healthcare
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Uterine Suction Device to Prevent Postpartum Hemorrhage in Scheduled Cesarean Delivery: A Pilot Study

Michael J. Brennan, George A. Saade, Neil Padharia Ray, Tetsuya Kawakita et al.
American Journal of Perinatology
Maternal and fetal healthcare
article

A Uterine Suction Device to Prevent Postpartum Hemorrhage in Scheduled Cesarean Delivery: A Pilot Study

Michael J. Brennan, George A. Saade, Neil Padharia Ray, Tetsuya Kawakita, Mark Rosen
article en

Abstract

Postpartum hemorrhage (PPH) remains a leading cause of maternal morbidity and mortality, with cesarean delivery posing a heightened risk. While interventions such as prophylactic tranexamic acid and balloon tamponade have limitations-especially when the cervix is not dilated-vacuum-assisted uterine tamponade may offer a novel intraoperative approach. This prospective pilot study evaluated the feasibility of the Daisy catheter, a cervical drain device developed by Raydiant Oximetry, Inc., designed to evacuate blood and promote uterine contraction through continuous negative-pressure suction.We enrolled ten pregnant individuals scheduled for cesarean delivery at a tertiary care center, all of whom had at least one PPH risk factor. Following fetal and placental delivery, the Daisy catheter was inserted trans-hysterotomically, advanced through the cervix, and connected to wall suction (-90 to -100 mm Hg) for 2 hours. Quantitative blood loss, perioperative hemoglobin change, ultrasound findings, and adverse events were recorded.Device placement succeeded in 9 of 10 cases; one failure was due to an undiagnosed cervical cerclage. Eight participants completed the full suction protocol. Mean hemoglobin decline from preoperative baseline to postoperative day 1 was 1.36 ± 0.47 g/dL, significantly lower than the 1.9 ± 1.1 g/dL observed in a historical cohort from 31 U.S. hospitals (<i>p</i> = 0.019). Ultrasound at 2 hours postpartum confirmed correct device placement, absence of intrauterine clot, and no evidence of trauma. Device removal was uncomplicated, and no adverse events were reported.These preliminary findings suggest that intraoperative use of the Daisy device is feasible, well-tolerated, and may reduce blood loss after cesarean delivery. Larger, randomized trials are warranted to evaluate its impact on transfusion rates, reoperation, and overall maternal outcomes, particularly in settings where alternative tamponade methods are limited. The ClinicalTrials.gov identifier is NCT06219538. · Vacuum tamponade used during cesarean delivery.. · Daisy device showed safe, feasible deployment.. · Hemoglobin drop was lower than the historical average..

American Journal of Perinatology
Eastern Virginia Medical School (US), Occidental Petroleum (United States) (US), Reproductive Science Center (US), Old Dominion University (US)
Openalex Percentile: Top 53%
Maternal and fetal healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.