A Discussion of Relevant and Meaningful Extraction Solvents for Extractables Testing of Implanted Medical Devices

During their clinical use, medical devices contact, directly or indirectly, tissue of the person (patient) whose medical condition is being treated or monitored by the device. During contact, substances present in or on the medical device can be leached from the device. Medical devices are chemically characterized to establish the patient's actual exposure to leachables.Establishing actual leachables from implanted methodical devices is problematic as so doing would require sampling of the <i>in vivo</i> environment in which the device is implanted. Thus, leachables from implanted medical devices are estimated by performing extraction studies. Often, such extraction studies use extraction solvents of multiple polarities to establish the device's total extractables profile (all possible extractables at their highest possible levels).As the potential range of extractables' polarities is large, using a small number of extraction solvents with widely different polarities can produce polarity gaps, where an extractable is either not revealed by the selected extraction solvents (an omission gap) or is underestimated in the extraction solvents (a magnitude gap). In either circumstance, the patient's exposure to extractables (as potential leachables) is underestimated.Furthermore, extraction studies which use extraction solvents with polarities outside of the polarity range of <i>in vivo</i> environments are inefficient, as they reveal extractables that cannot possibly be leachables at levels that cannot possibly be achieved during clinical use.After discussing these challenges from a thermodynamic and conceptual perspective, this author proposes that extraction studies for implanted medical devices use extraction solvents whose polarities bracket the polarity of the implanted <i>in vivo</i> environment and that a non-polar solvent never be used for bracketing purposes. To accomplish bracketing, this author introduces the polarity range termed seminon- polar" establishes the polarity of <i>in vivo</i> environments, and specifies polarity brackets.

Authors

Institutions

Publication Details

Journal
PDA Journal of Pharmaceutical Science and Technology
Published
2025-11-24
DOI
https://doi.org/10.5731/pdajpst.2025-000056.1
Primary Topic
Extraction and Separation Processes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Discussion of Relevant and Meaningful Extraction Solvents for Extractables Testing of Implanted Medical Devices

Dennis Jenke
PDA Journal of Pharmaceutical Science and Technology
Extraction and Separation Processes
article

A Discussion of Relevant and Meaningful Extraction Solvents for Extractables Testing of Implanted Medical Devices

Dennis Jenke
article en

Abstract

During their clinical use, medical devices contact, directly or indirectly, tissue of the person (patient) whose medical condition is being treated or monitored by the device. During contact, substances present in or on the medical device can be leached from the device. Medical devices are chemically characterized to establish the patient's actual exposure to leachables.Establishing actual leachables from implanted methodical devices is problematic as so doing would require sampling of the <i>in vivo</i> environment in which the device is implanted. Thus, leachables from implanted medical devices are estimated by performing extraction studies. Often, such extraction studies use extraction solvents of multiple polarities to establish the device's total extractables profile (all possible extractables at their highest possible levels).As the potential range of extractables' polarities is large, using a small number of extraction solvents with widely different polarities can produce polarity gaps, where an extractable is either not revealed by the selected extraction solvents (an omission gap) or is underestimated in the extraction solvents (a magnitude gap). In either circumstance, the patient's exposure to extractables (as potential leachables) is underestimated.Furthermore, extraction studies which use extraction solvents with polarities outside of the polarity range of <i>in vivo</i> environments are inefficient, as they reveal extractables that cannot possibly be leachables at levels that cannot possibly be achieved during clinical use.After discussing these challenges from a thermodynamic and conceptual perspective, this author proposes that extraction studies for implanted medical devices use extraction solvents whose polarities bracket the polarity of the implanted <i>in vivo</i> environment and that a non-polar solvent never be used for bracketing purposes. To accomplish bracketing, this author introduces the polarity range termed seminon- polar" establishes the polarity of <i>in vivo</i> environments, and specifies polarity brackets.

PDA Journal of Pharmaceutical Science and Technology
Scientific Solutions (United States) (US), Triad National Security (United States) (US)
Openalex Percentile: Top 63%
Extraction and Separation Processes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.