Evaluation of the small-field output factor in eclipse modeling methods using representative beam and measured data with averaged ionization chamber and diode detector measurements

Beam modeling for radiotherapy treatment planning systems (RTPS) can be performed using representative beam data (RBD) or direct measurements. However, RBD typically excludes output factor (OPF) measurements for fields smaller than 3 × 3 cm<sup>2</sup>. The Eclipse treatment planning system addresses this limitation by incorporating measured OPF data for fields as small as 1 × 1 cm<sup>2</sup>. Although existing studies have primarily examined the accuracy of small-field OPFs for plastic scintillator detectors, studies directly comparing the OPF values obtained through RBD modeling with and without OPF measurements for small field sizes are limited. Therefore, this study proposes a novel measurement approach using data averaged from an ion chamber and diode detector for small-field dosimetry to provide critical insights into the integration of OPFs for these small field sizes in RBD-based beam modeling. We systematically evaluated the impact of small-field OPF measurements on beam modeling accuracy by comparing three distinct approaches: (1) RBD-based modeling without small-field OPF data, (2) RBD-based modeling incorporating measured small-field OPF data, and (3) modeling based solely on measured data, with and without the inclusion of 1 × 1 cm<sup>2</sup> field sizes. In addition, we compared OPF values obtained from a W2 plastic scintillator detector with the averaged OPF values from a PinPoint 3D ion chamber and EDGE diode detector across multiple beam energies and flattening filter-free (FFF) configurations. Our analysis included field sizes ranging from 1 × 1 cm<sup>2</sup> to 40 × 40 cm<sup>2</sup>. The results demonstrated that for square fields, OPF calculation differences between RBD modeling with and without measured data were < 1.5%, < 4.5%, and < 4.5% at 1 × 1 cm<sup>2</sup>, and < 0.5%, < 1.5%, and < 1.5% at 2 × 2 cm<sup>2</sup>, respectively. The RBD group exhibited a trend in which the OPF difference increased with the expansion of the irradiation field size. Notably, the most significant variations between modeling approaches occurred along the upper jaw expansion direction in rectangular fields. This suggests that a thorough evaluation is necessary for modeling results with an OPF ≤ 1 × 1 cm<sup>2</sup>. This study highlights the advantages and disadvantages of beam modeling using measured OPF and RBD, providing valuable insights for future facilities that rely solely on RBD for beam modeling.

Authors

Institutions

Publication Details

Journal
Physical and Engineering Sciences in Medicine
Published
2025-11-24
DOI
https://doi.org/10.1007/s13246-025-01676-1
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the small-field output factor in eclipse modeling methods using representative beam and measured data with averaged ionization chamber and diode detector measurements

Yuki Kunii, Akira Nakamoto, Yoshinori Tanabe, Yūichi Sakamoto et al.
Physical and Engineering Sciences in Medicine
Advanced Radiotherapy Techniques
article

Evaluation of the small-field output factor in eclipse modeling methods using representative beam and measured data with averaged ionization chamber and diode detector measurements

Yuki Kunii, Akira Nakamoto, Yoshinori Tanabe, Yūichi Sakamoto, S. Takahashi, Kunio NISHIOKA
article en

Abstract

Beam modeling for radiotherapy treatment planning systems (RTPS) can be performed using representative beam data (RBD) or direct measurements. However, RBD typically excludes output factor (OPF) measurements for fields smaller than 3 × 3 cm<sup>2</sup>. The Eclipse treatment planning system addresses this limitation by incorporating measured OPF data for fields as small as 1 × 1 cm<sup>2</sup>. Although existing studies have primarily examined the accuracy of small-field OPFs for plastic scintillator detectors, studies directly comparing the OPF values obtained through RBD modeling with and without OPF measurements for small field sizes are limited. Therefore, this study proposes a novel measurement approach using data averaged from an ion chamber and diode detector for small-field dosimetry to provide critical insights into the integration of OPFs for these small field sizes in RBD-based beam modeling. We systematically evaluated the impact of small-field OPF measurements on beam modeling accuracy by comparing three distinct approaches: (1) RBD-based modeling without small-field OPF data, (2) RBD-based modeling incorporating measured small-field OPF data, and (3) modeling based solely on measured data, with and without the inclusion of 1 × 1 cm<sup>2</sup> field sizes. In addition, we compared OPF values obtained from a W2 plastic scintillator detector with the averaged OPF values from a PinPoint 3D ion chamber and EDGE diode detector across multiple beam energies and flattening filter-free (FFF) configurations. Our analysis included field sizes ranging from 1 × 1 cm<sup>2</sup> to 40 × 40 cm<sup>2</sup>. The results demonstrated that for square fields, OPF calculation differences between RBD modeling with and without measured data were < 1.5%, < 4.5%, and < 4.5% at 1 × 1 cm<sup>2</sup>, and < 0.5%, < 1.5%, and < 1.5% at 2 × 2 cm<sup>2</sup>, respectively. The RBD group exhibited a trend in which the OPF difference increased with the expansion of the irradiation field size. Notably, the most significant variations between modeling approaches occurred along the upper jaw expansion direction in rectangular fields. This suggests that a thorough evaluation is necessary for modeling results with an OPF ≤ 1 × 1 cm<sup>2</sup>. This study highlights the advantages and disadvantages of beam modeling using measured OPF and RBD, providing valuable insights for future facilities that rely solely on RBD for beam modeling.

Physical and Engineering Sciences in MedicineVol. 49(1)
Tokuyama (Japan) (JP), Okayama University (JP), Okayama University Hospital (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 55%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.