Accuracy and Reliability of an MR‐Compatible Dorsiflexion Ergometer for Dynamic 31P‐MRS: Comparison With a Clinical Dynamometer in Individuals With and Without Obesity
The clinical application of MR-compatible ergometers for muscle contractile assessment is limited by a lack of validation against standard clinical dynamometers. Moreover, the impact of obesity on the reliability of MR ergometer-based muscle contractile assessments and the quality of phosphorus-31 magnetic resonance spectroscopy (<sup>31</sup>P-MRS) data remains unclear. This study aimed to validate an MR-compatible ergometer against a clinical dynamometer and to evaluate the applicability of <sup>31</sup>P-MRS in individuals with severe obesity. Twenty adults (35-60 years) were recruited and divided into groups of nonobesity (BMI 18.5-30 kg/m<sup>2</sup>, n = 10) and severe obesity (BMI ≥ 35 kg/m<sup>2</sup>, n = 10), matched for age, sex, and height. Ankle dorsiflexion was assessed using both a clinical dynamometer and an MR ergometer, measuring maximal voluntary isometric contraction (MVIC) and a 4-min isotonic fatiguing exercise. <sup>31</sup>P-MRS was continuously acquired during the in-scanner exercise. Agreement between devices was assessed using Bland-Altman plots and intraclass correlation coefficients (ICCs). <sup>31</sup>P-MRS data quality was evaluated based on signal-to-noise ratio (SNR), uncertainty of fit (CRLB), and phosphocreatine (PCr) recovery fit (R<sup>2</sup>). Pearson's correlations examined relationships between muscle fatigue and metabolic parameters. All subjects successfully completed the protocol on both devices. The MR ergometer demonstrated moderate-to-excellent reliability (ICC ≥ 0.50) for most contractile parameters. While maximal torque, power, and work were underestimated on the MR ergometer (16-28%), this bias was consistent across BMI groups. <sup>31</sup>P-MRS met preset quality thresholds (SNR ≥ 5, CRLB < 20%, R<sup>2</sup> ≥ 0.70) in both groups. Dorsiflexion fatigue (reduction in power) correlated strongly (r ≥ 0.77) with metabolic changes, including PCr depletion (R<sup>2</sup> = 0.68), pH drop (R<sup>2</sup> = 0.59), PCr recovery time constant (R<sup>2</sup> = 0.62), and inorganic phosphate accumulation (Pi/PCr) (R<sup>2</sup> = 0.67). The MR ergometer demonstrated feasibility, acceptable reliability, and consistent <sup>31</sup>P-MRS data quality across BMI groups. These findings support the use of the MR ergometer for in-scanner dorsiflexor assessments, even in individuals with severe obesity.
Authors
- Gwenaël Layec (ORCID: https://orcid.org/0000-0002-5029-2457)
- Ole Emil Andersen
- Jack J. Miller (ORCID: https://orcid.org/0000-0002-6258-1299)
- Henning Boje Andersen (ORCID: https://orcid.org/0000-0002-4787-8502)
- Jens Meldgaard Bruun (ORCID: https://orcid.org/0000-0001-9937-5467)
- Jane A. Kent (ORCID: https://orcid.org/0000-0001-5487-2849)
- Anders Stouge (ORCID: https://orcid.org/0000-0001-8396-462X)
- Esben Søvsø Szocska Hansen (ORCID: https://orcid.org/0000-0001-5512-9870)
- Christoffer Laustsen (ORCID: https://orcid.org/0000-0002-0317-2911)
- Michael Væggemose (ORCID: https://orcid.org/0000-0001-7666-6819)
- Ladislav Valkovič (ORCID: https://orcid.org/0000-0003-2567-3642)
- Cuno Rasmussen
Institutions
- University of Nebraska at Omaha (US)
- Nebraska Medical Center (US)
- Steno Diabetes Centers (DK)
- Aarhus University (DK)
- Slovak Academy of Sciences (SK)
- University of Massachusetts Amherst (US)
- Aarhus University Hospital (DK)
- John Radcliffe Hospital (GB)
- University of Oxford (GB)
- Institute of Measurement Science of the Slovak Academy of Sciences (SK)
- Oxford BioMedica (United Kingdom) (GB)
- Aalborg University (DK)
Publication Details
- Journal
- NMR in Biomedicine
- Published
- 2025-11-23
- DOI
- https://doi.org/10.1002/nbm.70188
- Citations
- 1
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 1.36