False Lumen Haemodynamics in Type B Aortic Dissection: An in Vitro Study Using PIV and Patient-Specific Flexible Phantoms

Aortic dissection (AD) is a catastrophic vascular pathology caused by delamination of the vascular wall and the formation of a false lumen. False lumen haemodynamics is a key determinant of aneurysmal growth, rupture, and thrombosis. Quantifying the haemodynamics in the false lumen can provide markers to predict these events and stratify patient risk. While such metrics can be extracted from numerical simulations or imaging modalities such as 4D flow MR, high-resolution experimental data are needed to validate them. The present study provides an in vitro characterization of the flow inside the false lumen of a type B aortic dissection using a patient-specific flexible phantom and Particle Image Velocimetry. A mock circulatory loop imposing patient-specific flow waveforms at the inlet and outlets of the aortic phantom and a refractive index matching blood analog were employed. Time-resolved measurements of the velocity field in four selected planes of the false lumen were acquired. Compared against our previous work on the same dissection assuming rigid walls, the results demonstrate the impact of wall compliance on the flow in the false lumen. They revealed the generation of a jet during the systolic phase that enters the false lumen through the primary tear and impinging on the opposite wall with high velocity, generating a strong rotational flow therein. During the diastolic phase, a reversal of the flow was observed generating multiple vortical structures both inside the true and false lumen. Haemodynamic markers such as false lumen ejection fraction were calculated and compared with clinical measurements. The results provide an insight on AD haemodynamics and highlight the potential of this in vitro method as a validation tool for simulations, as well as to physically test interventions in vitro.

Authors

Institutions

Publication Details

Journal
Annals of Biomedical Engineering
Published
2025-11-22
DOI
https://doi.org/10.1007/s10439-025-03875-z
Primary Topic
Aortic Disease and Treatment Approaches
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

False Lumen Haemodynamics in Type B Aortic Dissection: An in Vitro Study Using PIV and Patient-Specific Flexible Phantoms

Vanessa Díaz‐Zuccarini, Shervanthi Homer‐Vanniasinkam, Stavroula Balabani, Anastasios Koulogiannis et al.
Annals of Biomedical Engineering
Aortic Disease and Treatment Approaches
article

False Lumen Haemodynamics in Type B Aortic Dissection: An in Vitro Study Using PIV and Patient-Specific Flexible Phantoms

Vanessa Díaz‐Zuccarini, Shervanthi Homer‐Vanniasinkam, Stavroula Balabani, Anastasios Koulogiannis, Qingfeng Li
article en

Abstract

Aortic dissection (AD) is a catastrophic vascular pathology caused by delamination of the vascular wall and the formation of a false lumen. False lumen haemodynamics is a key determinant of aneurysmal growth, rupture, and thrombosis. Quantifying the haemodynamics in the false lumen can provide markers to predict these events and stratify patient risk. While such metrics can be extracted from numerical simulations or imaging modalities such as 4D flow MR, high-resolution experimental data are needed to validate them. The present study provides an in vitro characterization of the flow inside the false lumen of a type B aortic dissection using a patient-specific flexible phantom and Particle Image Velocimetry. A mock circulatory loop imposing patient-specific flow waveforms at the inlet and outlets of the aortic phantom and a refractive index matching blood analog were employed. Time-resolved measurements of the velocity field in four selected planes of the false lumen were acquired. Compared against our previous work on the same dissection assuming rigid walls, the results demonstrate the impact of wall compliance on the flow in the false lumen. They revealed the generation of a jet during the systolic phase that enters the false lumen through the primary tear and impinging on the opposite wall with high velocity, generating a strong rotational flow therein. During the diastolic phase, a reversal of the flow was observed generating multiple vortical structures both inside the true and false lumen. Haemodynamic markers such as false lumen ejection fraction were calculated and compared with clinical measurements. The results provide an insight on AD haemodynamics and highlight the potential of this in vitro method as a validation tool for simulations, as well as to physically test interventions in vitro.

Annals of Biomedical EngineeringVol. 54(2)
Leeds Teaching Hospitals NHS Trust (GB), The London College (GB), University College London (GB)
Wellcome Trust, British Heart Foundation, University College London, Wellcome / EPSRC Centre for Interventional and Surgical Sciences, Directorate for Biological Sciences, Engineering and Physical Sciences Research Council, Biotechnology and Biological Sciences Research Council
Openalex Percentile: Top 49%
Aortic Disease and Treatment Approaches
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.