Dosimetric evaluation of Ethos 2.0 high‐fidelity mode for single‐isocenter SRS

The HFonRon template produced better quality plans, achieving significant improvements in normal tissue sparing (p < 0.0001), CI (p < 0.0001), and Falloff index (p < 0.0001) compared to other templates. All templates provided clinically acceptable target coverage. In the HFoffRoff configuration, 97.5 % of targets met the coverage criterion (V100% > 95%). The other setups achieved coverage in 99.2% (HFoffRon), 96.6% (HFonRoff), and 98.3% (HFonRon) of targets, respectively. Plan complexity decreased significantly (p < 0.0001) when enabling HF mode compared to both HF disabled templates CONCLUSIONS: His study shows that combining Ethos 2.0's high-fidelity mode with user-defined control rings yields superior single-isocenter SRS plans compared with planning strategies that omit one or both features. This approach improves dose conformity, reduces normal tissue exposure, and decreases plan complexity through semi-automated, template-based planning. These findings suggest that HF mode can be a valuable tool in clinical practice for optimizing SRS treatments for patients with brain metastases.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2025-11-21
DOI
https://doi.org/10.1002/acm2.70370
Primary Topic
Brain Metastases and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dosimetric evaluation of Ethos 2.0 high‐fidelity mode for single‐isocenter SRS

Joel A. Pogue, Dennis N. Stanley, Carlos E. Cardenas, Richard A. Popple et al.
Journal of Applied Clinical Medical Physics
Brain Metastases and Treatment
article

Dosimetric evaluation of Ethos 2.0 high‐fidelity mode for single‐isocenter SRS

Joel A. Pogue, Dennis N. Stanley, Carlos E. Cardenas, Richard A. Popple, Katherine A. Heinzman, Natalie Viscariello, Christopher B. Stanley, Yogesh Kumar, Udbhav S Ram, John B. Fiveash
article en

Abstract

The HFonRon template produced better quality plans, achieving significant improvements in normal tissue sparing (p < 0.0001), CI (p < 0.0001), and Falloff index (p < 0.0001) compared to other templates. All templates provided clinically acceptable target coverage. In the HFoffRoff configuration, 97.5 % of targets met the coverage criterion (V100% > 95%). The other setups achieved coverage in 99.2% (HFoffRon), 96.6% (HFonRoff), and 98.3% (HFonRon) of targets, respectively. Plan complexity decreased significantly (p < 0.0001) when enabling HF mode compared to both HF disabled templates CONCLUSIONS: His study shows that combining Ethos 2.0's high-fidelity mode with user-defined control rings yields superior single-isocenter SRS plans compared with planning strategies that omit one or both features. This approach improves dose conformity, reduces normal tissue exposure, and decreases plan complexity through semi-automated, template-based planning. These findings suggest that HF mode can be a valuable tool in clinical practice for optimizing SRS treatments for patients with brain metastases.

Journal of Applied Clinical Medical PhysicsVol. 26(12)
University of Alabama at Birmingham (US), McMaster University (CA)
Openalex Percentile: Top 49%
Brain Metastases and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Dosimetric evaluation of Ethos 2.0 high‐fidelity mode for single‐isocenter SRS — Joel A. Pogue, Dennis N. Stanley, et al. · Journal of Applied Clinical Medical Physics (2025) | TGRS Research Map | TGRS