Dosimetric evaluation of Ethos 2.0 high‐fidelity mode for single‐isocenter SRS
The HFonRon template produced better quality plans, achieving significant improvements in normal tissue sparing (p < 0.0001), CI (p < 0.0001), and Falloff index (p < 0.0001) compared to other templates. All templates provided clinically acceptable target coverage. In the HFoffRoff configuration, 97.5 % of targets met the coverage criterion (V100% > 95%). The other setups achieved coverage in 99.2% (HFoffRon), 96.6% (HFonRoff), and 98.3% (HFonRon) of targets, respectively. Plan complexity decreased significantly (p < 0.0001) when enabling HF mode compared to both HF disabled templates CONCLUSIONS: His study shows that combining Ethos 2.0's high-fidelity mode with user-defined control rings yields superior single-isocenter SRS plans compared with planning strategies that omit one or both features. This approach improves dose conformity, reduces normal tissue exposure, and decreases plan complexity through semi-automated, template-based planning. These findings suggest that HF mode can be a valuable tool in clinical practice for optimizing SRS treatments for patients with brain metastases.
Authors
- Joel A. Pogue (ORCID: https://orcid.org/0000-0003-4450-6698)
- Dennis N. Stanley (ORCID: https://orcid.org/0000-0003-4792-748X)
- Carlos E. Cardenas (ORCID: https://orcid.org/0000-0003-1414-3849)
- Richard A. Popple (ORCID: https://orcid.org/0000-0003-1252-738X)
- Katherine A. Heinzman
- Natalie Viscariello
- Christopher B. Stanley (ORCID: https://orcid.org/0000-0002-4226-7710)
- Yogesh Kumar (ORCID: https://orcid.org/0000-0002-2800-0235)
- Udbhav S Ram (ORCID: https://orcid.org/0000-0002-0457-7259)
- John B. Fiveash
Institutions
- University of Alabama at Birmingham (US)
- McMaster University (CA)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2025-11-21
- DOI
- https://doi.org/10.1002/acm2.70370
- Primary Topic
- Brain Metastases and Treatment
- Type
- article
- Field-Weighted Citation Impact
- 0.00