Physiological swelling imaging in human calf under stocking compression by sparse Bayesian learning implemented into electrical impedance tomography (SBL-EIT)

Physiological swelling in human calf has been imaged under stocking compression by the sparse Bayesian learning implemented into electrical impedance tomography (SBL-EIT) to evaluate the in situ treatment effect of various compression pressures. SBL-EIT reconstructs conductivity distribution [Formula: see text] to image excessive extracellular fluid in subcutaneous adipose tissue (SAT), indicating the susceptibilities to physiological swelling due to various compression pressures. The SBL-EIT was applied to the imaging of eight-subject calves during prolonged standing under three types of net compression pressures <i>P</i> <sup>net</sup> measured by pressure sensor - strong pressure: <i>P</i> <sup>net, Strong</sup> = 11.9 [Formula: see text] 2.0 mmHg, weak pressure: <i>P</i> <sup>net, Weak</sup> = 4.47 [Formula: see text] 3.1 mmHg, and control pressure: <i>P</i> <sup>net, Control</sup> = 0.00 [Formula: see text] 0.0 mmHg, respectively. From the experimental results, the spatial-mean conductivity ⟨<b>σ</b>⟩<sup>α2</sup> with two pre-processing steps to eliminate undesirable effects, i.e., the difference in skin condition and effect of wearing stockings itself, is the highest in the case of stocking with control pressure, followed by weak and strong pressures across all subjects. Moreover, the ⟨<b>σ</b>⟩<sup>α<i>2</i></sup> has a strong positive correlation with the conventional inversed impedance 1/<i>z</i> <sup>BIA</sup> by bioelectrical impedance analysis (BIA) (a correlation coefficient 0.528 < <i>R</i> < 0.990; <i>n</i> = 19 and <i>p</i> < 0.05), which is mainly increased during the prolonged standing. Moreover, various susceptibilities to physiological swelling are investigated based on the increase in [Formula: see text] for each subject, which is associated with subject external factors such as postural changes and circumference and internal factors like SAT.

Authors

Institutions

Publication Details

Journal
Biomedical Engineering Letters
Published
2025-11-22
DOI
https://doi.org/10.1007/s13534-025-00524-1
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Physiological swelling imaging in human calf under stocking compression by sparse Bayesian learning implemented into electrical impedance tomography (SBL-EIT)

Prima Asmara Sejati, Shinsuke Akita, Ryoma Ogawa, Masahiro Takei et al.
Biomedical Engineering Letters
Body Composition Measurement Techniques
article

Physiological swelling imaging in human calf under stocking compression by sparse Bayesian learning implemented into electrical impedance tomography (SBL-EIT)

Prima Asmara Sejati, Shinsuke Akita, Ryoma Ogawa, Masahiro Takei, Kota Asano
article en

Abstract

Physiological swelling in human calf has been imaged under stocking compression by the sparse Bayesian learning implemented into electrical impedance tomography (SBL-EIT) to evaluate the in situ treatment effect of various compression pressures. SBL-EIT reconstructs conductivity distribution [Formula: see text] to image excessive extracellular fluid in subcutaneous adipose tissue (SAT), indicating the susceptibilities to physiological swelling due to various compression pressures. The SBL-EIT was applied to the imaging of eight-subject calves during prolonged standing under three types of net compression pressures <i>P</i> <sup>net</sup> measured by pressure sensor - strong pressure: <i>P</i> <sup>net, Strong</sup> = 11.9 [Formula: see text] 2.0 mmHg, weak pressure: <i>P</i> <sup>net, Weak</sup> = 4.47 [Formula: see text] 3.1 mmHg, and control pressure: <i>P</i> <sup>net, Control</sup> = 0.00 [Formula: see text] 0.0 mmHg, respectively. From the experimental results, the spatial-mean conductivity ⟨<b>σ</b>⟩<sup>α2</sup> with two pre-processing steps to eliminate undesirable effects, i.e., the difference in skin condition and effect of wearing stockings itself, is the highest in the case of stocking with control pressure, followed by weak and strong pressures across all subjects. Moreover, the ⟨<b>σ</b>⟩<sup>α<i>2</i></sup> has a strong positive correlation with the conventional inversed impedance 1/<i>z</i> <sup>BIA</sup> by bioelectrical impedance analysis (BIA) (a correlation coefficient 0.528 < <i>R</i> < 0.990; <i>n</i> = 19 and <i>p</i> < 0.05), which is mainly increased during the prolonged standing. Moreover, various susceptibilities to physiological swelling are investigated based on the increase in [Formula: see text] for each subject, which is associated with subject external factors such as postural changes and circumference and internal factors like SAT.

Biomedical Engineering LettersVol. 16(1)
Chiba University (JP), Universitas Gadjah Mada (ID), Chiba University Hospital (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 56%
Body Composition Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.