Study of dose reduction in computed tomography using a time-of-flight scatter rejection mechanism in a 3D voxelized phantom
<i>Objective</i>. Scattered photons are the main source of contrast degradation in x-ray imaging when not filtered out appropriately. Measuring the time-of-flight (ToF) of x-ray photons has recently been revisited as an innovative way to discriminate between primary and scattered photons without significant loss in sensitivity in order to improve the contrast-to-noise ratio (CNR) in radiography and computed tomography (CT). The technique could allow for the removal of the anti-scatter grid (ASG) while still producing a CNR comparable to that of a fan-beam CT, allowing for potential dose reduction and more uniform axial spatial resolution when used in a cone-beam configuration.<i>Approach</i>. This work presents an evaluation study of the potential reduction in absorbed radiation dose in the body during a CT image acquisition when using time-correlated single photon counting CT in a cone-beam geometry. A comparative study is carried out using Monte Carlo simulations in GEANT4 Application for Tomographic Emission of a voxelized body of an adult female for a multi-slice fan beam CT scanner and a cone beam system using ToF measurements with various timing jitters or ASGs.<i>Main results</i>. For CNR equivalent torso images, simulation data in the high contrast region for ToF of 10, 50, 100, 200 ps full width at half maximum (FWHM) timing jitters show a dose reduction of 35%, 28%, 15%, 5% respectively compared to cone-beam CT with an ASG. Compared to a fan-beam setup without an ASG, the cone-beam ToF CT achieved a dose reduction of 13% and 4% for 10 and 50 ps FWHM timing jitter respectively, and an increase of 13% and 26% at 100 and 200 ps timing jitter. In the low contrast region, 10 ps ToF led to a reduction in dose of 15% and 50 ps led to an increase of 5%. Resolutions above 100 ps failed to accurately reject scattered photons and led to a considerable increase in dose.<i>Significance</i>. ToF scatter rejection shows potential for reducing radiation dose while preserving image quality when the timing resolution is 50 ps FWHM or better, but better ToF-based scatter rejection algorithms are required to enable this method for worse timing resolutions and low contrast regions.
Authors
- Julien Rossignol (ORCID: https://orcid.org/0000-0002-4760-5694)
- Yves Bérubé-Lauzière (ORCID: https://orcid.org/0000-0001-8106-7845)
- Réjean Fontaine (ORCID: https://orcid.org/0000-0002-5453-0168)
- Gabriel Bélanger (ORCID: https://orcid.org/0000-0003-1769-8399)
Institutions
- Université de Sherbrooke (CA)
- InterDigital (Canada) (CA)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2025-11-20
- DOI
- https://doi.org/10.1088/1361-6560/ae222f
- Primary Topic
- Digital Radiography and Breast Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00