Turning pan-sensor contact variability into a calibration tool for Sub-1% accuracy in heat capacity measurements with disk-type DSC

• Sub-1% accuracy in cp is achieved for small samples in disk-type DSC. • Pan-sensor contact variability is repurposed from an error source to a calibration tool. • Software-only correction uses isothermal heat flow as positioning metric. • Thermal model with leakage resistance explains the observed positioning effects. • Instrument-agnostic method applies to various materials and masses The accuracy of absolute heat capacity (c p ) measurements in disk-type differential scanning calorimetry (DSC) has been perennially compromised by a fundamental yet unresolved issue: the poor reproducibility of pan-sensor thermal contact resistance (PS-TCR) upon manual repositioning. This systematic error, frequently treated as random noise, routinely introduces significant uncertainties that limit the reliability of c p data, especially for small samples. This work introduces a methodological advancement to overcome this limitation. We demonstrate that the isothermal heat flow (Q̇ iso ), a readily measurable parameter in step-scan protocols, serves as a highly sensitive and quantitative signature of pan-positioning artifacts. By establishing a robust linear correlation between Q̇ iso and the integrated enthalpy signal (Q s ), we develop a novel correction procedure that effectively isolates and subtracts the positioning-dependent error component. This strategy, validated experimentally and supported by a refined thermal model, systematically reduces c p measurement uncertainty to below 1 % for sample masses as low as 5 mg, a level of precision previously unattainable in conventional disk-type DSCs without specialized equipment. Crucially, this approach requires no hardware modifications, relying solely on software-based post-processing of existing step-scan data. Our findings provide a practical and universally applicable solution to a decades-old problem, finally enabling high-accuracy c p determination on standard DSC instruments and democratizing precision for the broader materials science community.

Authors

Institutions

Publication Details

Journal
Thermochimica Acta
Published
2025-11-21
DOI
https://doi.org/10.1016/j.tca.2025.180192
Primary Topic
thermodynamics and calorimetric analyses
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Turning pan-sensor contact variability into a calibration tool for Sub-1% accuracy in heat capacity measurements with disk-type DSC

Frederic Hugenell, Yahya Rharbi
Thermochimica Acta
thermodynamics and calorimetric analyses
article

Turning pan-sensor contact variability into a calibration tool for Sub-1% accuracy in heat capacity measurements with disk-type DSC

Frederic Hugenell, Yahya Rharbi
article en

Abstract

• Sub-1% accuracy in cp is achieved for small samples in disk-type DSC. • Pan-sensor contact variability is repurposed from an error source to a calibration tool. • Software-only correction uses isothermal heat flow as positioning metric. • Thermal model with leakage resistance explains the observed positioning effects. • Instrument-agnostic method applies to various materials and masses The accuracy of absolute heat capacity (c p ) measurements in disk-type differential scanning calorimetry (DSC) has been perennially compromised by a fundamental yet unresolved issue: the poor reproducibility of pan-sensor thermal contact resistance (PS-TCR) upon manual repositioning. This systematic error, frequently treated as random noise, routinely introduces significant uncertainties that limit the reliability of c p data, especially for small samples. This work introduces a methodological advancement to overcome this limitation. We demonstrate that the isothermal heat flow (Q̇ iso ), a readily measurable parameter in step-scan protocols, serves as a highly sensitive and quantitative signature of pan-positioning artifacts. By establishing a robust linear correlation between Q̇ iso and the integrated enthalpy signal (Q s ), we develop a novel correction procedure that effectively isolates and subtracts the positioning-dependent error component. This strategy, validated experimentally and supported by a refined thermal model, systematically reduces c p measurement uncertainty to below 1 % for sample masses as low as 5 mg, a level of precision previously unattainable in conventional disk-type DSCs without specialized equipment. Crucially, this approach requires no hardware modifications, relying solely on software-based post-processing of existing step-scan data. Our findings provide a practical and universally applicable solution to a decades-old problem, finally enabling high-accuracy c p determination on standard DSC instruments and democratizing precision for the broader materials science community.

Thermochimica ActaVol. 755
Institut polytechnique de Grenoble (FR), Centre National de la Recherche Scientifique (FR), Université Grenoble Alpes (FR)
Agence Nationale de la Recherche, Institut Carnot PolyNat
Openalex Percentile: Top 68%
thermodynamics and calorimetric analyses
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.