<i>Objective</i>. Metallic implants introduce significant uncertainties in proton therapy by distorting computed tomography (CT) images and compromising dose calculations. This study evaluated single-energy CT (SECT), dual-energy CT (DECT) with iterative metal artifact reduction (IMAR) and megavoltage CT (MVCT) to optimize imaging and dosimetry during pencil beam scanning proton therapy.<i>Approach</i>. Controlled phantom studies were conducted using titanium and cobalt alloy spinal rods and a titanium plate embedded in solid water. Imaging protocols included SECT with MAR, DECT with IMAR and MVCT. DECT datasets were post-processed to generate virtual monoenergetic images (VMI), relative electron density (RED) and stopping power ratio (SPR) images. Geometric accuracy was assessed using the Dice similarity coefficient, the Hausdorff distance and the mean distance to agreement. CT number calibration was performed for each modality. Proton range and dose accuracy were validated using Gafchromic EBT3 films and ionization chamber arrays. Optimal protocols were applied to complex implants involving dental posts, a dental crown, spinal screws and hip prostheses. Treatment plans were generated in RayStation using robust optimization and range errors were compared between single field uniform dose (SFUD) and intensity modulated proton therapy (IMPT).<i>Main results</i>. MVCT achieved the highest geometric fidelity (Dice = 0.89, distance-to-agreement = 0.19 mm). SECT provided the most accurate density estimate for titanium (4.34 g cm<sup>-3</sup>) and RW3 (1.01 g cm<sup>-3</sup>). SPR or RED with IMAR yielded superior dose accuracy (γ>98%, dose deviation⩽1.1%, range error ≈ 1.1 mm). With complex implants, range errors varied between 0.8 mm (titanium stem) and 7.6 mm (cobalt screw head). SFUD plans demonstrated better range control than IMPT.<i>Significance</i>. SPR or RED, with IMAR, enabled accurate dose and range estimations for small titanium implants. For larger or high-<i>Z</i>implants, combining MVCT with SPR or RED, with IMAR, improved treatment precision due to their higher geometric and dosimetric accuracies, respectively.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-11-20
DOI
https://doi.org/10.1088/1361-6560/ae222d
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Computed tomography sequence integration for enhanced proton therapy in patients with metallic implants

G. Angellier, Baptiste Lhomel, J. Hérault, J. Feuillade et al.
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

Computed tomography sequence integration for enhanced proton therapy in patients with metallic implants

G. Angellier, Baptiste Lhomel, J. Hérault, J. Feuillade, Marie Vidal, Daniel C. Maneval, Pietro Mancosu, Benjamín Serrano, Régis Amblard, C. Colnard-Hofverberg, Luka McNeill, Ronan Penard
article en

Abstract

<i>Objective</i>. Metallic implants introduce significant uncertainties in proton therapy by distorting computed tomography (CT) images and compromising dose calculations. This study evaluated single-energy CT (SECT), dual-energy CT (DECT) with iterative metal artifact reduction (IMAR) and megavoltage CT (MVCT) to optimize imaging and dosimetry during pencil beam scanning proton therapy.<i>Approach</i>. Controlled phantom studies were conducted using titanium and cobalt alloy spinal rods and a titanium plate embedded in solid water. Imaging protocols included SECT with MAR, DECT with IMAR and MVCT. DECT datasets were post-processed to generate virtual monoenergetic images (VMI), relative electron density (RED) and stopping power ratio (SPR) images. Geometric accuracy was assessed using the Dice similarity coefficient, the Hausdorff distance and the mean distance to agreement. CT number calibration was performed for each modality. Proton range and dose accuracy were validated using Gafchromic EBT3 films and ionization chamber arrays. Optimal protocols were applied to complex implants involving dental posts, a dental crown, spinal screws and hip prostheses. Treatment plans were generated in RayStation using robust optimization and range errors were compared between single field uniform dose (SFUD) and intensity modulated proton therapy (IMPT).<i>Main results</i>. MVCT achieved the highest geometric fidelity (Dice = 0.89, distance-to-agreement = 0.19 mm). SECT provided the most accurate density estimate for titanium (4.34 g cm<sup>-3</sup>) and RW3 (1.01 g cm<sup>-3</sup>). SPR or RED with IMAR yielded superior dose accuracy (γ>98%, dose deviation⩽1.1%, range error ≈ 1.1 mm). With complex implants, range errors varied between 0.8 mm (titanium stem) and 7.6 mm (cobalt screw head). SFUD plans demonstrated better range control than IMPT.<i>Significance</i>. SPR or RED, with IMAR, enabled accurate dose and range estimations for small titanium implants. For larger or high-<i>Z</i>implants, combining MVCT with SPR or RED, with IMAR, improved treatment precision due to their higher geometric and dosimetric accuracies, respectively.

Physics in Medicine and BiologyVol. 70(23)
Humanitas University (IT), Princess Grace Hospital Centre (MC), Centre Antoine Lacassagne (FR), Fondazione Humanitas per la Ricerca (IT)
Openalex Percentile: Top 65%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.