Accuracy and Reduced Sleep Disruption of a Wearable Smartwatch for Nocturnal Blood Pressure Monitoring: A Validation Study

Nocturnal blood pressure is a significant predictor of cardiovascular risk, but traditional ambulatory blood pressure monitoring (ABPM) can disrupt sleep. This study aimed to validate the accuracy of the Huawei Watch D2 (WD2), a wrist-worn oscillometric device, and assess its impact on sleep quality compared to conventional ABPM. This prospective, single-center trial first validated the WD2 against a mercury sphygmomanometer in 85 participants according to ANSI/AAMI/ISO 81060-2:2018 standards in both seated and supine positions. Subsequently, the device's nocturnal accuracy and sleep impact were compared against a traditional ABPM device in 46 participants over a three-night protocol. The WD2 met all ISO accuracy standards. During nighttime monitoring, there was no significant difference between the WD2 and ABPM for mean systolic or diastolic blood pressure. However, the number of awakenings was significantly lower on nights with the WD2 alone compared to nights with the ABPM device (p = 0.016). In conclusion, the Huawei Watch D2 is a clinically validated device that provides nocturnal blood pressure readings comparable to traditional ABPM with the significant advantage of minimal sleep disruption, positioning it as a valuable alternative for nocturnal hypertension monitoring.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Hypertension
Published
2025-11-01
DOI
https://doi.org/10.1111/jch.70183
Primary Topic
Sleep and related disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy and Reduced Sleep Disruption of a Wearable Smartwatch for Nocturnal Blood Pressure Monitoring: A Validation Study

Zhuang Tian, Lu Yang, Beibei Wang, Yanbo Liu et al.
Journal of Clinical Hypertension
Sleep and related disorders
article

Accuracy and Reduced Sleep Disruption of a Wearable Smartwatch for Nocturnal Blood Pressure Monitoring: A Validation Study

Zhuang Tian, Lu Yang, Beibei Wang, Yanbo Liu, Jiabo Wu, T Zhang
article en

Abstract

Nocturnal blood pressure is a significant predictor of cardiovascular risk, but traditional ambulatory blood pressure monitoring (ABPM) can disrupt sleep. This study aimed to validate the accuracy of the Huawei Watch D2 (WD2), a wrist-worn oscillometric device, and assess its impact on sleep quality compared to conventional ABPM. This prospective, single-center trial first validated the WD2 against a mercury sphygmomanometer in 85 participants according to ANSI/AAMI/ISO 81060-2:2018 standards in both seated and supine positions. Subsequently, the device's nocturnal accuracy and sleep impact were compared against a traditional ABPM device in 46 participants over a three-night protocol. The WD2 met all ISO accuracy standards. During nighttime monitoring, there was no significant difference between the WD2 and ABPM for mean systolic or diastolic blood pressure. However, the number of awakenings was significantly lower on nights with the WD2 alone compared to nights with the ABPM device (p = 0.016). In conclusion, the Huawei Watch D2 is a clinically validated device that provides nocturnal blood pressure readings comparable to traditional ABPM with the significant advantage of minimal sleep disruption, positioning it as a valuable alternative for nocturnal hypertension monitoring.

Journal of Clinical HypertensionVol. 27(11)
Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Peking Union Medical College Hospital (CN)
Peking Union Medical College Hospital, Peking Union Medical College
Openalex Percentile: Top 62%
Sleep and related disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Accuracy and Reduced Sleep Disruption of a Wearable Smartwatch for Nocturnal Blood Pressure Monitoring: A Validation Study — Zhuang Tian, Lu Yang, et al. · Journal of Clinical Hypertension (2025) | TGRS Research Map | TGRS