Added value of a wrist-worn device for assessing tremor in Parkinson’s disease: reliability and validity of tremor evaluation at home

Clinical assessment and self-report of tremor in Parkinson's disease (PD) have several limitations, including recall bias and lack of sensitivity. Wearable-based technology offers an opportunity to address these shortcomings. We investigated the test-retest reliability of wearable tremor assessment and compared the different approaches of assessing tremor. We analyzed data from 219 participants of the ProPark study. In 27% of people with PD, self-report and clinical observation disagreed about the presence of tremor. Wearable-based tremor assessment had excellent test-retest reliability after 3 days of data recording (ICC(2,3) lower bound of 95%CI > 0.90 for all tremor metrics). Wearable-derived tremor amplitude, duration, and volume were significantly associated with clinical assessment of tremor (all p < 0.001), and tremor volume was consistent with patient self-report of tremor presence when clinical observation failed to detect tremor (p = 0.040). In conclusion, wearable sensors can provide accurate and relevant information about the presence and severity of tremor in PD.

Authors

Institutions

Publication Details

Journal
npj Parkinson s Disease
Published
2025-11-20
DOI
https://doi.org/10.1038/s41531-025-01163-0
Primary Topic
Neurological disorders and treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Added value of a wrist-worn device for assessing tremor in Parkinson’s disease: reliability and validity of tremor evaluation at home

Dagmar H. Hepp, R. H. A. Weijer, Vasileios Exadaktylos, Kenneth Ocran et al.
npj Parkinson s Disease
Neurological disorders and treatments
article

Added value of a wrist-worn device for assessing tremor in Parkinson’s disease: reliability and validity of tremor evaluation at home

Dagmar H. Hepp, R. H. A. Weijer, Vasileios Exadaktylos, Kenneth Ocran, Antónia Szántó, J.J. van Hilten
article en

Abstract

Clinical assessment and self-report of tremor in Parkinson's disease (PD) have several limitations, including recall bias and lack of sensitivity. Wearable-based technology offers an opportunity to address these shortcomings. We investigated the test-retest reliability of wearable tremor assessment and compared the different approaches of assessing tremor. We analyzed data from 219 participants of the ProPark study. In 27% of people with PD, self-report and clinical observation disagreed about the presence of tremor. Wearable-based tremor assessment had excellent test-retest reliability after 3 days of data recording (ICC(2,3) lower bound of 95%CI > 0.90 for all tremor metrics). Wearable-derived tremor amplitude, duration, and volume were significantly associated with clinical assessment of tremor (all p < 0.001), and tremor volume was consistent with patient self-report of tremor presence when clinical observation failed to detect tremor (p = 0.040). In conclusion, wearable sensors can provide accurate and relevant information about the presence and severity of tremor in PD.

npj Parkinson s DiseaseVol. 11(1)
Rijksmuseum (NL), New York Academy of Medicine (US), Leiden University Medical Center (NL), Centre for Human Drug Research (NL), Amsterdam University Medical Centers (NL)
ZonMw, Hersenstichting
Openalex Percentile: Top 51%
Neurological disorders and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.