In-vitro degradation and corrosion protection of chitosan composite coatings containing hydroxyapatite or bioactive glass on Ti6Al4V implants: A comparative study

This research systematically evaluates the biodegradation and corrosion behavior of a novel class of composite coatings for bone implants based on chitosan (CS) and two bioactive components (Hydroxyapatite (HA) and Bioactive glass (BG)) in simulated physiological conditions through a comparative approach. Chitosan (CS)-based composite coatings (HACS and BGCS) were deposited on Ti6Al4V substrates via electrophoretic deposition (EPD) using unified processing parameters. Initially, the primary materials, relevant suspensions, and final coatings were studied and compared by scanning electron microscopy (SEM), energy-dispersive spectroscopy (EDS), Zetasizer, and X-ray diffractometer (XRD) methods. While positive ZPs for both HACS (55.44 ± 2.43 mV) and BGCS (35.62 ± 2.95 mV) suspensions were aligned with cathodic deposition of coatings, a large difference between recorded current densities during EPD for HACS (6.98 ± 0.38 mA·cm −2 ) and BGCS (56.14 ± 3.99 mA·cm −2 ) were noteworthy. The in-vitro degradation of BGCS and HACS coatings in DMEM solution was assessed not only under a conventional static condition but also within a fluid-dynamic degradation setup. The mass and pH evolution, surface morphology, compositional changes, and release of metallic ions were investigated over the degradation period by gravimetric, optical microscope, SEM, FTIR, and ICP-MS techniques, respectively. While the majority of BGCS degradation (above 3 mg·cm −2 ) occurred only after 8 h of immersion, a maximum mass loss of about 2 mg·cm −2 happened for HACS coatings on day 10 of the test. Electrochemical corrosion studies of coated samples compared to plain substrates in DMEM solution were carried out using open circuit potential (OCP), potentiodynamic polarization (PDP), and electrochemical impedance spectroscopy (EIS) tests, and the equivalent circuit was defined to model the experimental data. Interestingly, the corrosion rate of the uncoated sample (9.13E−05 mm·year −1 ) was about three times and ten times less than HACS and BGCS samples, respectively. It was revealed that the surface pretreatment (polishing) method can have a decisive effect on the corrosion protection behavior of Ti6Al4V samples, coated with CS-based EPD coatings. • Biodegradation and corrosion behavior of chitosan-based coatings evaluated in DMEM • The rate of degradation in BG-containing coatings significantly higher than of HA counterparts • Surface pretreatment affects Ti6Al4V corrosion resistance and coating performance. • Particle morphology, charge, and solubility affect EPD coating structure.

Authors

Institutions

Publication Details

Journal
Surface and Coatings Technology
Published
2025-11-20
DOI
https://doi.org/10.1016/j.surfcoat.2025.132962
Primary Topic
Bone Tissue Engineering Materials
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In-vitro degradation and corrosion protection of chitosan composite coatings containing hydroxyapatite or bioactive glass on Ti6Al4V implants: A comparative study

Muhammad Saqib, Aldo Roberto Boccaccini, Jörg Opitz, Natalia Beshchasna et al.
Surface and Coatings Technology
Bone Tissue Engineering Materials
article

In-vitro degradation and corrosion protection of chitosan composite coatings containing hydroxyapatite or bioactive glass on Ti6Al4V implants: A comparative study

Muhammad Saqib, Aldo Roberto Boccaccini, Jörg Opitz, Natalia Beshchasna, Ivan Lopez Carrasco, Anton Ficai, Rezvan Azari
article en

Abstract

This research systematically evaluates the biodegradation and corrosion behavior of a novel class of composite coatings for bone implants based on chitosan (CS) and two bioactive components (Hydroxyapatite (HA) and Bioactive glass (BG)) in simulated physiological conditions through a comparative approach. Chitosan (CS)-based composite coatings (HACS and BGCS) were deposited on Ti6Al4V substrates via electrophoretic deposition (EPD) using unified processing parameters. Initially, the primary materials, relevant suspensions, and final coatings were studied and compared by scanning electron microscopy (SEM), energy-dispersive spectroscopy (EDS), Zetasizer, and X-ray diffractometer (XRD) methods. While positive ZPs for both HACS (55.44 ± 2.43 mV) and BGCS (35.62 ± 2.95 mV) suspensions were aligned with cathodic deposition of coatings, a large difference between recorded current densities during EPD for HACS (6.98 ± 0.38 mA·cm −2 ) and BGCS (56.14 ± 3.99 mA·cm −2 ) were noteworthy. The in-vitro degradation of BGCS and HACS coatings in DMEM solution was assessed not only under a conventional static condition but also within a fluid-dynamic degradation setup. The mass and pH evolution, surface morphology, compositional changes, and release of metallic ions were investigated over the degradation period by gravimetric, optical microscope, SEM, FTIR, and ICP-MS techniques, respectively. While the majority of BGCS degradation (above 3 mg·cm −2 ) occurred only after 8 h of immersion, a maximum mass loss of about 2 mg·cm −2 happened for HACS coatings on day 10 of the test. Electrochemical corrosion studies of coated samples compared to plain substrates in DMEM solution were carried out using open circuit potential (OCP), potentiodynamic polarization (PDP), and electrochemical impedance spectroscopy (EIS) tests, and the equivalent circuit was defined to model the experimental data. Interestingly, the corrosion rate of the uncoated sample (9.13E−05 mm·year −1 ) was about three times and ten times less than HACS and BGCS samples, respectively. It was revealed that the surface pretreatment (polishing) method can have a decisive effect on the corrosion protection behavior of Ti6Al4V samples, coated with CS-based EPD coatings. • Biodegradation and corrosion behavior of chitosan-based coatings evaluated in DMEM • The rate of degradation in BG-containing coatings significantly higher than of HA counterparts • Surface pretreatment affects Ti6Al4V corrosion resistance and coating performance. • Particle morphology, charge, and solubility affect EPD coating structure.

Surface and Coatings TechnologyVol. 519
Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Academia Oamenilor de Știință din România (RO), Fraunhofer Institute for Ceramic Technologies and Systems (DE), Universitatea Națională de Știință și Tehnologie Politehnica București (RO)
Openalex Percentile: Top 66%
Bone Tissue Engineering Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.