Mitigating electrolyte measurement discrepancies in high triglycerides samples: a comparative analysis of direct and indirect ISE for sodium, potassium and chloride measurements

Elevated triglyceride (TG) levels, common in hypertriglyceridemia, can significantly interfere with electrolyte analysis, particularly by the indirect ion-selective electrode (ISE) method. However, comprehensive data on the concentration-dependent measurement differences for potassium, sodium and chloride, along with validated corrective algorithms, are lacking. This study assessed the discrepancies in these electrolytes between direct and indirect ISE methods in serum samples with high TGs, while developing correction formulas for the indirect ISE. A total of 154 serum samples with high TGs and 50 control serum samples were analyzed in this retrospective cross-sectional study. Triglycerides were measured using colorimetric methods on the Roche Cobas 8000 analyzer (Roche Laboratories, Basel, Switzerland). Sodium, potassium and chloride were measured with direct ISE (Vitros 5600 Integrated System; Ortho-Clinical Diagnostics, Inc., Raritan, NJ) and indirect ISE (Roche Cobas 8000 analyzer). Our results revealed significant negative biases in the indirect ISE, particularly in samples with TG >20.00 mmol/L. For TG levels between 20.01 and 30.00 mmol/L, the negative biases for sodium, potassium and chloride were -2.31%, -3.86% and -4.58%, respectively. Notably, in the subgroup with TG >60.00 mmol/L, the negative biases reached their maximum values: -12.05% for potassium, -6.88% for sodium, and -10.59% for chloride. Additionally, linear correction formulas that aligned indirect results with direct measurements were developed and validated. Post-correction, differences fell within clinical thresholds (Diff_Na of |4| mmol/L, Diff_Cl of |4| mmol/L, Diff_K of |0.5| mmol/L). Collectively, high TGs impact electrolyte measurements by indirect ISE, but the correction formulas might mitigate the discrepancies. These correction formulas were platform-specific, and their generalizability to other analytical systems requires further investigation.

Authors

Institutions

Publication Details

Journal
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2025-10-03
DOI
https://doi.org/10.1080/00365513.2025.2585468
Primary Topic
Electrolyte and hormonal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mitigating electrolyte measurement discrepancies in high triglycerides samples: a comparative analysis of direct and indirect ISE for sodium, potassium and chloride measurements

Zhixue Wang, Junqun Liao, Xie Chang-Li, Yulan Hou
Scandinavian Journal of Clinical and Laboratory Investigation
Electrolyte and hormonal disorders
article

Mitigating electrolyte measurement discrepancies in high triglycerides samples: a comparative analysis of direct and indirect ISE for sodium, potassium and chloride measurements

Zhixue Wang, Junqun Liao, Xie Chang-Li, Yulan Hou
article en

Abstract

Elevated triglyceride (TG) levels, common in hypertriglyceridemia, can significantly interfere with electrolyte analysis, particularly by the indirect ion-selective electrode (ISE) method. However, comprehensive data on the concentration-dependent measurement differences for potassium, sodium and chloride, along with validated corrective algorithms, are lacking. This study assessed the discrepancies in these electrolytes between direct and indirect ISE methods in serum samples with high TGs, while developing correction formulas for the indirect ISE. A total of 154 serum samples with high TGs and 50 control serum samples were analyzed in this retrospective cross-sectional study. Triglycerides were measured using colorimetric methods on the Roche Cobas 8000 analyzer (Roche Laboratories, Basel, Switzerland). Sodium, potassium and chloride were measured with direct ISE (Vitros 5600 Integrated System; Ortho-Clinical Diagnostics, Inc., Raritan, NJ) and indirect ISE (Roche Cobas 8000 analyzer). Our results revealed significant negative biases in the indirect ISE, particularly in samples with TG >20.00 mmol/L. For TG levels between 20.01 and 30.00 mmol/L, the negative biases for sodium, potassium and chloride were -2.31%, -3.86% and -4.58%, respectively. Notably, in the subgroup with TG >60.00 mmol/L, the negative biases reached their maximum values: -12.05% for potassium, -6.88% for sodium, and -10.59% for chloride. Additionally, linear correction formulas that aligned indirect results with direct measurements were developed and validated. Post-correction, differences fell within clinical thresholds (Diff_Na of |4| mmol/L, Diff_Cl of |4| mmol/L, Diff_K of |0.5| mmol/L). Collectively, high TGs impact electrolyte measurements by indirect ISE, but the correction formulas might mitigate the discrepancies. These correction formulas were platform-specific, and their generalizability to other analytical systems requires further investigation.

Scandinavian Journal of Clinical and Laboratory InvestigationVol. 85(7)
People's Hospital of Bishan District (CN), The Affiliated Yongchuan Hospital of Chongqing Medical University (CN), Chongqing Medical University (CN)
Openalex Percentile: Top 50%
Electrolyte and hormonal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.