Characterisation of the Timepix4 silicon sensor detector for low energy X-ray applications

This study aims to provide a characterisation of the Timepix4 hybrid pixel detector (HPD) in terms of energy calibration, energy resolution, temperature response and angular response for low-energy photon applications (20–150 keV). The Timepix4 detector used in this study was bump-bonded to a 300 μ m thick silicon sensor. The characterisation was performed using X-ray fluorescence lines from several materials and 3 radioactive sources: 55 Fe, 109 Cd and 241 Am. In this context, the Timepix4 HPD shows excellent performances, with an energy resolution ranging from 43.52% at 5.89 keV to 7.84% at 59.54 keV. The measured energy resolution was slightly better than with previous versions of the detector. A strong angular dependence was observed, with a photon count ratio decreasing by up to a factor of 2 for 55 Fe and 85% for 241 Am with a 90 ° rotation. The detector should therefore be used with restricted angular ranges to limit incorrect photonic fluences. Temperature was observed over the tested range to have a minimal impact. The relative energy shift for the 241 Am 59.54 keV peak varied between 0.74% from 10 °C to 1% at 40 °C, showing a linear relationship between the temperature variation and the position of the photopeak. No correlation between the temperature variation and the number of counts in the photopeak was observed, but a small energy resolution loss with the increasing temperature was identified. The use of correction factors or the implementation of the associated uncertainty should be considered in the case of large temperature variations in between measurements. • Timepix4 is the latest version of the Timepix ASIC. • Timepix4 demonstrated strong spectroscopic performances. • Temperature (10-40 °C) minimally affects Timepix4 energy and fluence accuracy. • Timepix4 has limited angular response for low-energy X-rays.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2025-11-16
DOI
https://doi.org/10.1016/j.nima.2025.171163
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterisation of the Timepix4 silicon sensor detector for low energy X-ray applications

X. Llopart, J. Alozy, François Bochud, L. Tlustos et al.
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Particle Detector Development and Performance
article

Characterisation of the Timepix4 silicon sensor detector for low energy X-ray applications

X. Llopart, J. Alozy, François Bochud, L. Tlustos, Pierre Carbonez, M. Campbell, J. Damet, T. Genetay, L. Desorgher
article en

Abstract

This study aims to provide a characterisation of the Timepix4 hybrid pixel detector (HPD) in terms of energy calibration, energy resolution, temperature response and angular response for low-energy photon applications (20–150 keV). The Timepix4 detector used in this study was bump-bonded to a 300 μ m thick silicon sensor. The characterisation was performed using X-ray fluorescence lines from several materials and 3 radioactive sources: 55 Fe, 109 Cd and 241 Am. In this context, the Timepix4 HPD shows excellent performances, with an energy resolution ranging from 43.52% at 5.89 keV to 7.84% at 59.54 keV. The measured energy resolution was slightly better than with previous versions of the detector. A strong angular dependence was observed, with a photon count ratio decreasing by up to a factor of 2 for 55 Fe and 85% for 241 Am with a 90 ° rotation. The detector should therefore be used with restricted angular ranges to limit incorrect photonic fluences. Temperature was observed over the tested range to have a minimal impact. The relative energy shift for the 241 Am 59.54 keV peak varied between 0.74% from 10 °C to 1% at 40 °C, showing a linear relationship between the temperature variation and the position of the photopeak. No correlation between the temperature variation and the number of counts in the photopeak was observed, but a small energy resolution loss with the increasing temperature was identified. The use of correction factors or the implementation of the associated uncertainty should be considered in the case of large temperature variations in between measurements. • Timepix4 is the latest version of the Timepix ASIC. • Timepix4 demonstrated strong spectroscopic performances. • Temperature (10-40 °C) minimally affects Timepix4 energy and fluence accuracy. • Timepix4 has limited angular response for low-energy X-rays.

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1083
Czech Technical University in Prague (CZ), European Organization for Nuclear Research (CH), University of Otago (NZ), University of Lausanne (CH)
CERN
Openalex Percentile: Top 84%
Particle Detector Development and Performance
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.