Preliminary analysis of the clinical safety of cochlear implants in non-implanted humans under magnetic resonance-specific conditions

In non-implanted volunteers with the CS-30A externally affixed at the anatomical site, MRI at both 1.5 and 3.0 T produced no reports of pain, electrical-like sensations, or thermal burns. Pre- and post-scan position assessments confirmed the absence of displacement, and comprehensive functional and magnet performance testing demonstrated stable operation throughout. These device-specific findings, obtained under the tested MRI parameters, provide preliminary but important evidence supporting the compatibility of the CS-30A with MRI.

Authors

Institutions

Publication Details

Journal
Cochlear Implants International
Published
2025-09-03
DOI
https://doi.org/10.1080/14670100.2025.2581944
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Preliminary analysis of the clinical safety of cochlear implants in non-implanted humans under magnetic resonance-specific conditions

Siqi Gao, Kaiyue Zhao, D.-Y. Yang, Mengxue Zhang et al.
Cochlear Implants International
Advanced MRI Techniques and Applications
article

Preliminary analysis of the clinical safety of cochlear implants in non-implanted humans under magnetic resonance-specific conditions

Siqi Gao, Kaiyue Zhao, D.-Y. Yang, Mengxue Zhang, Jiaying Wang, Zheng Qin Yin, Li Yan Yuan, Qin Li, Jiayu Ni
article en

Abstract

In non-implanted volunteers with the CS-30A externally affixed at the anatomical site, MRI at both 1.5 and 3.0 T produced no reports of pain, electrical-like sensations, or thermal burns. Pre- and post-scan position assessments confirmed the absence of displacement, and comprehensive functional and magnet performance testing demonstrated stable operation throughout. These device-specific findings, obtained under the tested MRI parameters, provide preliminary but important evidence supporting the compatibility of the CS-30A with MRI.

Cochlear Implants InternationalVol. 26(5)
Hangzhou Normal University (CN), Affiliated Hospital of Hangzhou Normal University (CN), Third People 's Hospital of Chongqing (CN), Beike Biotechnology (China) (CN)
Openalex Percentile: Top 50%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.