Uracil in plasma: comparison of two in-house-developed LC-MS/MS methods
For the analysis of uracil, both presented methods, which are based on simplified sample preparation procedures, provide accurate results without significantly impacting the clinical interpretation. However, the results indicate a potential gray area of analytical imprecision from 14 to 20 μg/L, encompassing the lower clinical decision limit (16 μg/L).