Evaluation of a universal talus implant during gait: a combined musculoskeletal and finite element modelling approach

Universal talus implant has emerged as an innovative solution for talus bone collapse, aiming to retain the clinical benefits of custom total talus replacement while addressing its logistical drawbacks. A subject-specific combined musculoskeletal-finite element (MSK-FE) modeling framework was developed to evaluate two universal talus implant designs during dynamic gait: a purely cobalt chromium (CoCr) implant, and an implant coated with polycarbonate-urethane (PCU), both compared to the native talus. To do so, a MSK simulation of the stance phase of gait was conducted to estimate joint kinematics and joint reaction forces in the ankle complex, with a subsequent dynamic FE simulation performed to assess contact characteristics in terms of contact area and pressure, in cartilages surrounding the talus/implant. The FE model was built directly from the bone geometries of the MSK model to ensure consistency across the study. Results showed that the PCU-coated implant more closely replicated native biomechanics, while the CoCr implant produced consistently higher pressures and smaller contact regions. Normalized RMSE across gait confirmed lower deviation from the native case for the PCU-implant in most joints. These findings highlight the potential of PCU coated implants in improving contact mechanics in articular cartilage as well as the potential of the universal implant topology. This is the first study to dynamically evaluate intra-articular behaviour in all joints surrounding the talus bone during gait, and particularly by analysing the performance of universal talus implants, demonstrating the utility of a MSK-FE approach and offering valuable insights into implant performance under physiological conditions, informing future implant design.

Authors

Institutions

Publication Details

Journal
Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials
Published
2025-11-17
DOI
https://doi.org/10.1016/j.jmbbm.2025.107278
Primary Topic
Foot and Ankle Surgery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of a universal talus implant during gait: a combined musculoskeletal and finite element modelling approach

Marwan El‐Rich, Stephen J. Ferguson, A. Hafez, Tao Liu et al.
Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials
Foot and Ankle Surgery
article

Evaluation of a universal talus implant during gait: a combined musculoskeletal and finite element modelling approach

Marwan El‐Rich, Stephen J. Ferguson, A. Hafez, Tao Liu, Sami Al Shweiki, Naod T Mogos
article en

Abstract

Universal talus implant has emerged as an innovative solution for talus bone collapse, aiming to retain the clinical benefits of custom total talus replacement while addressing its logistical drawbacks. A subject-specific combined musculoskeletal-finite element (MSK-FE) modeling framework was developed to evaluate two universal talus implant designs during dynamic gait: a purely cobalt chromium (CoCr) implant, and an implant coated with polycarbonate-urethane (PCU), both compared to the native talus. To do so, a MSK simulation of the stance phase of gait was conducted to estimate joint kinematics and joint reaction forces in the ankle complex, with a subsequent dynamic FE simulation performed to assess contact characteristics in terms of contact area and pressure, in cartilages surrounding the talus/implant. The FE model was built directly from the bone geometries of the MSK model to ensure consistency across the study. Results showed that the PCU-coated implant more closely replicated native biomechanics, while the CoCr implant produced consistently higher pressures and smaller contact regions. Normalized RMSE across gait confirmed lower deviation from the native case for the PCU-implant in most joints. These findings highlight the potential of PCU coated implants in improving contact mechanics in articular cartilage as well as the potential of the universal implant topology. This is the first study to dynamically evaluate intra-articular behaviour in all joints surrounding the talus bone during gait, and particularly by analysing the performance of universal talus implants, demonstrating the utility of a MSK-FE approach and offering valuable insights into implant performance under physiological conditions, informing future implant design.

Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materialsVol. 174
Khalifa University of Science and Technology (AE), ETH Zurich (CH), Ontario Tech University (CA), Institute for Biomedical Engineering (CH)
Openalex Percentile: Top 56%
Foot and Ankle Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.