In the diagnosis of venous thromboembolism (VTE), the use of point-of-care (POC) D-dimer testing is emerging as a promising alternative to laboratory-based D-dimer assays. These POC devices, easier to use, raise the question of their analytical and clinical performances. We evaluated the analytical performances of the POC Pathfast D-dimer chemiluminescence assay and compared its results with two laboratory-based D-dimer assays: Vidas D-dimer Exclusion II and STA-Liatest D-Di Plus. Within- and between-run imprecision were evaluated using low-and high-level quality control plasma samples and healthy controls and patients’ whole blood. Limit of quantification and linearity were determined according to the Clinical and Laboratory Standards Institute (CLSI). Comparisons were performed using the Passing-Bablok and Bland-Altman analysis, using fresh citrated whole blood and plasma samples collected from 150 consecutive routine samples. The within- and between-run imprecision of the assay did not exceed 4.4% and 5.8%, respectively. Our study found higher D-Dimer results with the Pathfast D-dimer assay (whole blood) than with the Vidas D-dimer Exclusion II and STA-Liatest D-Di Plus assays (plasma) with mean bias of 1.16 and 1.20 μg/mL FEU, respectively. Around the most clinically relevant area (<1 μg/mL FEU), bias was smallest, from 0.07 to 0.15 μg/mL FEU. The Pathfast D-dimer assay is a rapid POC test for determination of D-dimer with good analytical performance. However, these results may be further investigated by a well-designed prospective clinical evaluation in a context of VTE suspicion and appropriate scoring. • The use of point-of-care (POC) D-dimer testing is emerging as a promising alternative to laboratory-based D-dimer assays. • These POC devices raise the question of their analytical and clinical performances and should be evaluated. • The Pathfast D-dimer assay is a rapid POC test for determination of D-dimer with good analytical performance. • These results may be further investigated by a well-designed prospective clinical evaluation in a context of VTE suspicion and appropriate scoring.

Authors

Institutions

Publication Details

Journal
Thrombosis Update
Published
2025-11-17
DOI
https://doi.org/10.1016/j.tru.2025.100228
Primary Topic
Venous Thromboembolism Diagnosis and Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical performance of the quantitative point-of-care PATHFAST D-dimer assay

Virginie Fourneyron, Alexandre Trapani, M. Tillier, Bruno Pereira et al.
Thrombosis Update
Venous Thromboembolism Diagnosis and Management
article

Analytical performance of the quantitative point-of-care PATHFAST D-dimer assay

Virginie Fourneyron, Alexandre Trapani, M. Tillier, Bruno Pereira, Thomas Sinegre, S. Senectaire, A.F. Sapin, Laurie Talon, Jonathan Douxfils, Aurélien Lebreton, A. Trayaud, T. Dureng
article en

Abstract

In the diagnosis of venous thromboembolism (VTE), the use of point-of-care (POC) D-dimer testing is emerging as a promising alternative to laboratory-based D-dimer assays. These POC devices, easier to use, raise the question of their analytical and clinical performances. We evaluated the analytical performances of the POC Pathfast D-dimer chemiluminescence assay and compared its results with two laboratory-based D-dimer assays: Vidas D-dimer Exclusion II and STA-Liatest D-Di Plus. Within- and between-run imprecision were evaluated using low-and high-level quality control plasma samples and healthy controls and patients’ whole blood. Limit of quantification and linearity were determined according to the Clinical and Laboratory Standards Institute (CLSI). Comparisons were performed using the Passing-Bablok and Bland-Altman analysis, using fresh citrated whole blood and plasma samples collected from 150 consecutive routine samples. The within- and between-run imprecision of the assay did not exceed 4.4% and 5.8%, respectively. Our study found higher D-Dimer results with the Pathfast D-dimer assay (whole blood) than with the Vidas D-dimer Exclusion II and STA-Liatest D-Di Plus assays (plasma) with mean bias of 1.16 and 1.20 μg/mL FEU, respectively. Around the most clinically relevant area (<1 μg/mL FEU), bias was smallest, from 0.07 to 0.15 μg/mL FEU. The Pathfast D-dimer assay is a rapid POC test for determination of D-dimer with good analytical performance. However, these results may be further investigated by a well-designed prospective clinical evaluation in a context of VTE suspicion and appropriate scoring. • The use of point-of-care (POC) D-dimer testing is emerging as a promising alternative to laboratory-based D-dimer assays. • These POC devices raise the question of their analytical and clinical performances and should be evaluated. • The Pathfast D-dimer assay is a rapid POC test for determination of D-dimer with good analytical performance. • These results may be further investigated by a well-designed prospective clinical evaluation in a context of VTE suspicion and appropriate scoring.

Thrombosis UpdateVol. 21
University of Namur (BE), Institut National de Recherche pour l'Agriculture, l'Alimentation et l'Environnement (FR), Centre Hospitalier Universitaire de Clermont-Ferrand (FR), Centre de Recherche en Nutrition Humaine d'Auvergne (FR)
Openalex Percentile: Top 45%
Venous Thromboembolism Diagnosis and Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.