Cytokine adsorption properties of a polymethylmethacrylate (PMMA) hollow-fiber membrane: an in vitro study

Abstract Background The aims of this study were to analyze the relationship between the membrane surface area and cytokine adsorption performance of polymethylmethacrylate (PMMA) hemofilters as well as the distribution of cytokines inside the membrane after contact with a cytokine. Methods Briefly, 1 L of bovine plasma with interleukin (IL)-6 was circulated through four types of PMMA hemofilters with different membrane surface areas (0.3, 0.6, 1.3, and 1.8 m 2 ) without dialysis and filtration, and changes in the IL-6 concentration in the plasma were observed. In addition, mini-modules comprising PMMA hollow fibers were circulated in an IL-6 solution for 24 h, and the distribution of IL-6 in the PMMA membrane was inferred by detecting the secondary ion image of 12 C 14 N − , which reflects nitrogen-containing organic compounds (in this case, IL-6), using a nanoscale secondary ion mass spectrometer (NanoSIMS). Results After 24 h of circulation, the IL-6 removal rates for hemofilters with membrane surface areas of 0.3, 0.6, 1.3, and 1.8 m 2 were 74%, 84%, 89%, and 89%, respectively. The times required to reduce the initial concentration by 50% were 486 ± 25 min, 175 ± 16 min, 104 ± 6.7 min, and 88 ± 20 min, respectively ( p < 0.01). Using a 1.8 m 2 hemofilter, in vitro circulation tests for IL-8, IL-10, and tumor necrosis factor (TNF)- α showed removal rates of 87%, 85%, and 54%, respectively, at 4 h. The observed 12 C 14 N − signal indicates a possible distribution of IL-6 across the PMMA hollow-fiber membrane from the blood side to the dialysate side after the IL-6 solution was circulated through the mini-module for 24 h. Conclusion PMMA hemofilters rapidly adsorbed IL-6 in an in vitro circulation test, depending on the membrane surface area. IL-8, IL-10, and TNF- α were also removed. IL-6 was inferred to permeate the PMMA membrane and be widely adsorbed throughout the membrane.

Authors

Institutions

Publication Details

Journal
Renal Replacement Therapy
Published
2025-11-15
DOI
https://doi.org/10.1186/s41100-025-00687-8
Primary Topic
Membrane Separation Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Cytokine adsorption properties of a polymethylmethacrylate (PMMA) hollow-fiber membrane: an in vitro study

Yukiko Hayashi, M. Fujita, Wataru Hara, Yuri Watanabe et al.
Renal Replacement Therapy
Membrane Separation Technologies
article

Cytokine adsorption properties of a polymethylmethacrylate (PMMA) hollow-fiber membrane: an in vitro study

Yukiko Hayashi, M. Fujita, Wataru Hara, Yuri Watanabe, Sari Sugaya, Kazuhiro Matsuda, Kensaku Ohashi
article en

Abstract

Abstract Background The aims of this study were to analyze the relationship between the membrane surface area and cytokine adsorption performance of polymethylmethacrylate (PMMA) hemofilters as well as the distribution of cytokines inside the membrane after contact with a cytokine. Methods Briefly, 1 L of bovine plasma with interleukin (IL)-6 was circulated through four types of PMMA hemofilters with different membrane surface areas (0.3, 0.6, 1.3, and 1.8 m 2 ) without dialysis and filtration, and changes in the IL-6 concentration in the plasma were observed. In addition, mini-modules comprising PMMA hollow fibers were circulated in an IL-6 solution for 24 h, and the distribution of IL-6 in the PMMA membrane was inferred by detecting the secondary ion image of 12 C 14 N − , which reflects nitrogen-containing organic compounds (in this case, IL-6), using a nanoscale secondary ion mass spectrometer (NanoSIMS). Results After 24 h of circulation, the IL-6 removal rates for hemofilters with membrane surface areas of 0.3, 0.6, 1.3, and 1.8 m 2 were 74%, 84%, 89%, and 89%, respectively. The times required to reduce the initial concentration by 50% were 486 ± 25 min, 175 ± 16 min, 104 ± 6.7 min, and 88 ± 20 min, respectively ( p < 0.01). Using a 1.8 m 2 hemofilter, in vitro circulation tests for IL-8, IL-10, and tumor necrosis factor (TNF)- α showed removal rates of 87%, 85%, and 54%, respectively, at 4 h. The observed 12 C 14 N − signal indicates a possible distribution of IL-6 across the PMMA hollow-fiber membrane from the blood side to the dialysate side after the IL-6 solution was circulated through the mini-module for 24 h. Conclusion PMMA hemofilters rapidly adsorbed IL-6 in an in vitro circulation test, depending on the membrane surface area. IL-8, IL-10, and TNF- α were also removed. IL-6 was inferred to permeate the PMMA membrane and be widely adsorbed throughout the membrane.

Renal Replacement TherapyVol. 11(1)
Toray (United States) (US)
Openalex Percentile: Top 68%
Membrane Separation Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.