Effect of the Modified Look–Locker Inversion Recovery Sampling Scheme on Accuracy and Heart Rate Variability in Myocardial T1 Mapping

This study aimed to clarify the accuracy and variability of heart rate variability effects on T1 relaxation time values acquired using different sampling schemes of the modified Look-Locker inversion recovery (MOLLI) method. Five MOLLI sampling schemes-5(3)3, 5s(3s)3s, 4(1)3(1)2, 4s(1s)3s(1s)2s, and 3(3)3(3)5-were compared using a simulated electrocardiogram signal in a phantom experiment. The 5(3)3 and 5s(3s)3s schemes demonstrated high accuracy across the range of native and contrast-enhanced T1 values, with correlation coefficients exceeding 0.95 compared with the reference T1. Regarding heart rate variability, the 5(3)3 scheme showed minimal variation in the measured value across different heart rates, regardless of the T1 value, with an error rate below 6% relative to the reference T1. Therefore, the 5(3)3 scheme can achieve highly accurate measurements over a wide T1 range and is robust against heart rate variability.

Authors

Institutions

Publication Details

Journal
Magnetic Resonance in Medical Sciences
Published
2025-11-13
DOI
https://doi.org/10.2463/mrms.tn.2025-0022
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of the Modified Look–Locker Inversion Recovery Sampling Scheme on Accuracy and Heart Rate Variability in Myocardial T1 Mapping

Keita Fukushima, Yuta Endo, Kenichi Yokoyama, Akihito Nakanishi
Magnetic Resonance in Medical Sciences
Cardiac Imaging and Diagnostics
article

Effect of the Modified Look–Locker Inversion Recovery Sampling Scheme on Accuracy and Heart Rate Variability in Myocardial T1 Mapping

Keita Fukushima, Yuta Endo, Kenichi Yokoyama, Akihito Nakanishi
article en

Abstract

This study aimed to clarify the accuracy and variability of heart rate variability effects on T1 relaxation time values acquired using different sampling schemes of the modified Look-Locker inversion recovery (MOLLI) method. Five MOLLI sampling schemes-5(3)3, 5s(3s)3s, 4(1)3(1)2, 4s(1s)3s(1s)2s, and 3(3)3(3)5-were compared using a simulated electrocardiogram signal in a phantom experiment. The 5(3)3 and 5s(3s)3s schemes demonstrated high accuracy across the range of native and contrast-enhanced T1 values, with correlation coefficients exceeding 0.95 compared with the reference T1. Regarding heart rate variability, the 5(3)3 scheme showed minimal variation in the measured value across different heart rates, regardless of the T1 value, with an error rate below 6% relative to the reference T1. Therefore, the 5(3)3 scheme can achieve highly accurate measurements over a wide T1 range and is robust against heart rate variability.

Magnetic Resonance in Medical SciencesVol. 25(1)
Kyorin University (JP), Kyorin University Hospital (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 51%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.