Implementation of a one‐tube flow cytometry panel for measurable residual disease detection in multiple myeloma patients in clinical routine

Multiparametric flow cytometry is a highly valuable method for the assessment of measurable residual disease (MRD) in multiple myeloma patients. The aim of the study was to evaluate a one-tube MRD panel on a DxFlex flow cytometer including all EuroFlow recommended immunophenotypic markers (i.e., cytoplasmic light chain kappa and lambda, CD19, CD27, CD38, CD45, CD56, CD81, CD117, CD138) extended by CD200 to a total of 11 fluorochromes in one tube. Bone marrow aspirates from clinical routine underwent an ammonium chloride-based bulk lysis, followed by staining the surface antibodies and, after permeabilization, kappa, and lambda light-chain intracellular staining. We acquired 1 × 10<sup>7</sup> cells per sample with a DxFlex flow cytometer. We determined the limit of detection (LOD) and lower limit of quantification (LLOQ) as recommended by the International Myeloma Working Group. For the clinical evaluation, 68 samples from 53 patients with multiple myeloma under or after treatment were analyzed with the one-tube MRD panel, and the results were compared with our routine plasma cell panel (RPCP). Six of the 68 samples were additionally sent to another laboratory to confirm our results with the contemporary EuroFlow next-generation flow (NGF) assay. For our novel one-tube MRD panel we determined a LOD of 0.00016% (1.6 × 10<sup>-6</sup>) and a LLOQ of 0.00059% (5.9 × 10<sup>-6</sup>). Out of 68 specimens, 55 (80.9%) showed concordant results between the MRD- and the RPC-panel. Thirteen (19.1%) specimens showed a distinct population of abnormal plasma cells with the MRD panel not detectable with the RPC panel. The six samples simultaneously measured with our novel one-tube MRD panel and the EuroFlow NGF assay showed concordant results. The novel one-tube MRD panel meets the quality specifications of the International Myeloma Working Group. Our clinical evaluation found higher sensitivity for the one-tube MRD panel when compared to our RPC panel and concordant results with the contemporary EuroFlow NGF assay. Therefore, our novel one-tube MRD panel is well-suited for detecting MRD in multiple myeloma patients.

Authors

Institutions

Publication Details

Journal
Cytometry Part B Clinical Cytometry
Published
2025-11-14
DOI
https://doi.org/10.1002/cyto.b.22264
Primary Topic
Multiple Myeloma Research and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Implementation of a one‐tube flow cytometry panel for measurable residual disease detection in multiple myeloma patients in clinical routine

Benjamin Dieplinger, Sigrid Machherndl‐Spandl, Katrin Hefler‐Frischmuth, Irene Straßl et al.
Cytometry Part B Clinical Cytometry
Multiple Myeloma Research and Treatments
article

Implementation of a one‐tube flow cytometry panel for measurable residual disease detection in multiple myeloma patients in clinical routine

Benjamin Dieplinger, Sigrid Machherndl‐Spandl, Katrin Hefler‐Frischmuth, Irene Straßl, Johann Gruber, Peter Bettelheim, Isabella Herbring, Karin Deutsch‐Biedermann, Christine Kimbacher
article en

Abstract

Multiparametric flow cytometry is a highly valuable method for the assessment of measurable residual disease (MRD) in multiple myeloma patients. The aim of the study was to evaluate a one-tube MRD panel on a DxFlex flow cytometer including all EuroFlow recommended immunophenotypic markers (i.e., cytoplasmic light chain kappa and lambda, CD19, CD27, CD38, CD45, CD56, CD81, CD117, CD138) extended by CD200 to a total of 11 fluorochromes in one tube. Bone marrow aspirates from clinical routine underwent an ammonium chloride-based bulk lysis, followed by staining the surface antibodies and, after permeabilization, kappa, and lambda light-chain intracellular staining. We acquired 1 × 10<sup>7</sup> cells per sample with a DxFlex flow cytometer. We determined the limit of detection (LOD) and lower limit of quantification (LLOQ) as recommended by the International Myeloma Working Group. For the clinical evaluation, 68 samples from 53 patients with multiple myeloma under or after treatment were analyzed with the one-tube MRD panel, and the results were compared with our routine plasma cell panel (RPCP). Six of the 68 samples were additionally sent to another laboratory to confirm our results with the contemporary EuroFlow next-generation flow (NGF) assay. For our novel one-tube MRD panel we determined a LOD of 0.00016% (1.6 × 10<sup>-6</sup>) and a LLOQ of 0.00059% (5.9 × 10<sup>-6</sup>). Out of 68 specimens, 55 (80.9%) showed concordant results between the MRD- and the RPC-panel. Thirteen (19.1%) specimens showed a distinct population of abnormal plasma cells with the MRD panel not detectable with the RPC panel. The six samples simultaneously measured with our novel one-tube MRD panel and the EuroFlow NGF assay showed concordant results. The novel one-tube MRD panel meets the quality specifications of the International Myeloma Working Group. Our clinical evaluation found higher sensitivity for the one-tube MRD panel when compared to our RPC panel and concordant results with the contemporary EuroFlow NGF assay. Therefore, our novel one-tube MRD panel is well-suited for detecting MRD in multiple myeloma patients.

Cytometry Part B Clinical CytometryVol. 110(1)
Convent Hospital of the Brothers of Saint John of God (AT), Krankenhaus der Elisabethinen (AT)
Openalex Percentile: Top 55%
Multiple Myeloma Research and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.