Design Optimization and Mechanical Performance Evaluation of a Modified Coronary IV-OCT Catheter Adapted for Intracranial Navigation: A Preclinical Study

The application of intravascular optical coherence tomography (IV-OCT) in neurovascular interventions is constrained by the mechanical inadequacy of conventional catheters in navigating the complex intracranial vasculature. To address this, we modified a coronary IV-OCT catheter, enhancing its mechanical performance for neurovascular applications. The modified catheter featured a 300 mm over-the-microwire segment and a dual-structured shaft (distal 50 mm nonbraided, proximal 250 mm braided) to improve trackability and pushability. We compared the modified and conventional catheters using a benchtop model with a simulated vessel path and an in vivo swine model. Trackability and pushability were quantitatively measured using resistance (N) and advancement distance (mm) in the simulated path. In the animal model, indirect performance metrics included the catheter tension angle (CTA) and pass of catheter (POC) through the fourth curvature of the external carotid artery (ECA). The modified catheter demonstrated superior pushability (172.9 ± 1.96 mm vs. 127.9 ± 2.86 mm, <i>p</i> < 0.05) and increased resistance (1.47 ± 0.036 N vs. 0.69 ± 0.032 N, <i>p</i> < 0.05). In vivo analysis further showed a significantly greater CTA (115.8 ± 8.5° vs. 77.6 ± 10.3°, <i>p</i> < 0.05) and higher POC success rate (83.3% vs. 11.1%, <i>p</i> < 0.05). These results indicate that the modified coronary IV-OCT catheter offers enhanced mechanical performance, suggesting its potential for safe and effective use in neurovascular procedures.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2025-11-12
DOI
https://doi.org/10.3390/bios15110755
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design Optimization and Mechanical Performance Evaluation of a Modified Coronary IV-OCT Catheter Adapted for Intracranial Navigation: A Preclinical Study

Joonho Chung, Min-Yong Kwon, Hyuck‐Jun Yoon, Sang‐Hyun An et al.
Biosensors
Coronary Interventions and Diagnostics
article

Design Optimization and Mechanical Performance Evaluation of a Modified Coronary IV-OCT Catheter Adapted for Intracranial Navigation: A Preclinical Study

Joonho Chung, Min-Yong Kwon, Hyuck‐Jun Yoon, Sang‐Hyun An, Seung‐Ho Hur, Jae Hyun Kim, Sae Min Kwon, Jun-Sik Kim, Tahsin Nairuz, Woo Jin Kim, Jong‐Ha Lee, Chang‐Hyun Kim, Young-Suk Hwang
article en

Abstract

The application of intravascular optical coherence tomography (IV-OCT) in neurovascular interventions is constrained by the mechanical inadequacy of conventional catheters in navigating the complex intracranial vasculature. To address this, we modified a coronary IV-OCT catheter, enhancing its mechanical performance for neurovascular applications. The modified catheter featured a 300 mm over-the-microwire segment and a dual-structured shaft (distal 50 mm nonbraided, proximal 250 mm braided) to improve trackability and pushability. We compared the modified and conventional catheters using a benchtop model with a simulated vessel path and an in vivo swine model. Trackability and pushability were quantitatively measured using resistance (N) and advancement distance (mm) in the simulated path. In the animal model, indirect performance metrics included the catheter tension angle (CTA) and pass of catheter (POC) through the fourth curvature of the external carotid artery (ECA). The modified catheter demonstrated superior pushability (172.9 ± 1.96 mm vs. 127.9 ± 2.86 mm, <i>p</i> < 0.05) and increased resistance (1.47 ± 0.036 N vs. 0.69 ± 0.032 N, <i>p</i> < 0.05). In vivo analysis further showed a significantly greater CTA (115.8 ± 8.5° vs. 77.6 ± 10.3°, <i>p</i> < 0.05) and higher POC success rate (83.3% vs. 11.1%, <i>p</i> < 0.05). These results indicate that the modified coronary IV-OCT catheter offers enhanced mechanical performance, suggesting its potential for safe and effective use in neurovascular procedures.

BiosensorsVol. 15(11)
Yonsei University (KR), Daegu-Gyeongbuk Medical Innovation Foundation (KR), Keimyung University (KR), Korea Institute of Industrial Technology (KR)
Korea Basic Science Institute
Openalex Percentile: Top 47%
Coronary Interventions and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.