Mechanical Performance of Two <scp>NiTi</scp> Rotary Files: Effects of Alloy and Design of Fatigue, Torsion, Bending, and Fracture Behavior

This study evaluated the mechanical performance of nickel-titanium (NiTi) endodontic instruments with identical tip diameters but different alloy compositions and designs. The study compared two geometric configurations of WaveOne Gold instruments (Respectively, WO I and WO II) manufactured using three different NiTi alloys: SE-Wire (non-heat-treated superelastic NiTi), M-Wire, and gold alloy. WO I instruments have a variable taper (25/0.08) and dual convex triangular cross-section, while WO II instruments feature a reduced taper (25/0.07), square cross-section, and shorter handle. Six groups were tested for cyclic fatigue resistance, torsional strength, bending behavior, deformation, and compressive load. Fracture patterns were examined via scanning electron microscopy (SEM). A two-way ANOVA evaluated interactions between alloy type and instrument design. Alloy heat treatment and design significantly affected performance. Gold alloy instruments showed superior flexibility and the highest time-to-fracture (TTF) values, particularly in WO II. WO II files required less pressure to navigate canals and demonstrated better bending adaptability due to their reduced taper and square cross-section. In contrast, SE-Wire instruments had the lowest fatigue resistance and highest pressure loads, indicating their austenitic phase and lack of thermal treatment. WO I instruments had higher torsional strength but fractured at lower angular deflections. SEM revealed distinct fracture mechanisms based on alloy and generation. WO II instruments outperformed WO I in flexibility, cyclic fatigue resistance, and canal tracking, particularly with thermomechanically treated alloys like Gold. The relationship between alloy composition and instrument design is crucial for optimizing performance and minimizing procedural risks.

Authors

Institutions

Publication Details

Journal
Microscopy Research and Technique
Published
2025-11-14
DOI
https://doi.org/10.1002/jemt.70093
Primary Topic
Endodontics and Root Canal Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mechanical Performance of Two NiTi Rotary Files: Effects of Alloy and Design of Fatigue, Torsion, Bending, and Fracture Behavior

Giulia Malvicini, Vincenzo Tosco, José Aranguren, Alejandro R. Pérez et al.
Microscopy Research and Technique
Endodontics and Root Canal Treatments
article

Mechanical Performance of Two NiTi Rotary Files: Effects of Alloy and Design of Fatigue, Torsion, Bending, and Fracture Behavior

Giulia Malvicini, Vincenzo Tosco, José Aranguren, Alejandro R. Pérez, C.F. Bonilla, Gema Galvan
article en

Abstract

This study evaluated the mechanical performance of nickel-titanium (NiTi) endodontic instruments with identical tip diameters but different alloy compositions and designs. The study compared two geometric configurations of WaveOne Gold instruments (Respectively, WO I and WO II) manufactured using three different NiTi alloys: SE-Wire (non-heat-treated superelastic NiTi), M-Wire, and gold alloy. WO I instruments have a variable taper (25/0.08) and dual convex triangular cross-section, while WO II instruments feature a reduced taper (25/0.07), square cross-section, and shorter handle. Six groups were tested for cyclic fatigue resistance, torsional strength, bending behavior, deformation, and compressive load. Fracture patterns were examined via scanning electron microscopy (SEM). A two-way ANOVA evaluated interactions between alloy type and instrument design. Alloy heat treatment and design significantly affected performance. Gold alloy instruments showed superior flexibility and the highest time-to-fracture (TTF) values, particularly in WO II. WO II files required less pressure to navigate canals and demonstrated better bending adaptability due to their reduced taper and square cross-section. In contrast, SE-Wire instruments had the lowest fatigue resistance and highest pressure loads, indicating their austenitic phase and lack of thermal treatment. WO I instruments had higher torsional strength but fractured at lower angular deflections. SEM revealed distinct fracture mechanisms based on alloy and generation. WO II instruments outperformed WO I in flexibility, cyclic fatigue resistance, and canal tracking, particularly with thermomechanically treated alloys like Gold. The relationship between alloy composition and instrument design is crucial for optimizing performance and minimizing procedural risks.

Microscopy Research and TechniqueVol. 89(4)
University of Siena (IT), Marche Polytechnic University (IT), Universidad Rey Juan Carlos (ES), Nova Southeastern University (US)
Openalex Percentile: Top 51%
Endodontics and Root Canal Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.