Comparing pulse oximeter performance using a common functional tester versus controlled desaturation studies on healthy participants

Functional testers are designed to evaluate select pulse oximeter characteristics but are often misused to validate device accuracy, potentially providing false reassurance. This study evaluated whether the Fluke ProSim8 (FPS8) could accurately predict oximeter performance during human controlled desaturation studies or identify performance differences under low signal conditions. 12 oximeters were tested using two FPS8 protocols: (1) an 'SpO₂ plateau' protocol which mimicked controlled desaturation studies by evaluating device performance over a range of simulated SpO<sub>2</sub> (70-100%), and (2) a 'signal space' protocol designed to assess device accuracy under varying modulation and transmission conditions. Each device also underwent controlled desaturation testing in healthy adults. Six of the 12 oximeters passed (ARMS ≤ 3%) the SpO₂ plateau protocol; however, three of these failed (ARMS > 3%) human testing. At lower simulated saturations, most devices overestimated SpO₂. In the signal space protocol, oximeters performed well under high signal conditions, but many failed to produce readings or showed SpO₂ errors > 3% under low signal conditions. On average, oximeters failed to generate a reading 20.2 ± 7.2 times out of 60 attempts. Ten devices passed (ARMS < 3%) the signal space protocol, but two of these failed human testing. Oximeter performance on the FPS8 did not correlate with human performance (R² = 0.08 for the plateau protocol; R² = 0.01 for the signal space protocol). The FPS8 did not reliably predict oximeter accuracy in human desaturation studies or under low signal conditions; current functional tester protocols are limited in predicting real-world oximeter performance.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Monitoring and Computing
Published
2025-11-14
DOI
https://doi.org/10.1007/s10877-025-01381-0
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing pulse oximeter performance using a common functional tester versus controlled desaturation studies on healthy participants

Michael A. Bernstein, Isabella Auchus, Ronald Bisegerwa, Tyler J. Law et al.
Journal of Clinical Monitoring and Computing
Non-Invasive Vital Sign Monitoring
article

Comparing pulse oximeter performance using a common functional tester versus controlled desaturation studies on healthy participants

Michael A. Bernstein, Isabella Auchus, Ronald Bisegerwa, Tyler J. Law, Rocío Zamora, Michael S. Lipnick, Philip E. Bickler, Shamsudini Hashi, Jake Dove, John Feiner, Seif B. Elmankabadi, Gregory Leeb, Danni Chen, Yu Chou, Lily Ortiz, Ella Behnke, Fekir Negussie
article en

Abstract

Functional testers are designed to evaluate select pulse oximeter characteristics but are often misused to validate device accuracy, potentially providing false reassurance. This study evaluated whether the Fluke ProSim8 (FPS8) could accurately predict oximeter performance during human controlled desaturation studies or identify performance differences under low signal conditions. 12 oximeters were tested using two FPS8 protocols: (1) an 'SpO₂ plateau' protocol which mimicked controlled desaturation studies by evaluating device performance over a range of simulated SpO<sub>2</sub> (70-100%), and (2) a 'signal space' protocol designed to assess device accuracy under varying modulation and transmission conditions. Each device also underwent controlled desaturation testing in healthy adults. Six of the 12 oximeters passed (ARMS ≤ 3%) the SpO₂ plateau protocol; however, three of these failed (ARMS > 3%) human testing. At lower simulated saturations, most devices overestimated SpO₂. In the signal space protocol, oximeters performed well under high signal conditions, but many failed to produce readings or showed SpO₂ errors > 3% under low signal conditions. On average, oximeters failed to generate a reading 20.2 ± 7.2 times out of 60 attempts. Ten devices passed (ARMS < 3%) the signal space protocol, but two of these failed human testing. Oximeter performance on the FPS8 did not correlate with human performance (R² = 0.08 for the plateau protocol; R² = 0.01 for the signal space protocol). The FPS8 did not reliably predict oximeter accuracy in human desaturation studies or under low signal conditions; current functional tester protocols are limited in predicting real-world oximeter performance.

Journal of Clinical Monitoring and Computing
San Francisco General Hospital (US), University of California, San Francisco (US), Medtronic (United States) (US), Makerere University (UG), University of San Francisco (US)
Gordon and Betty Moore Foundation, University of California, San Francisco
Openalex Percentile: Top 67%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.