Method Development, Validation and Stability Study of Diltiazem by RP-HPLC Method
Background: The main goal of this experiment was to create and test a reliable and fast HPLC method for measuring the amount of Diltiazem, a drug that lowers blood pressure, in large quantities and to conduct stability studies. Methods: For the method development we use column of octa decyl silane which was bonded to porous silica with particle size 5µ ODS C-18 (250×4.6mm, 5µm), methanol and water mixture v/v proportion (90:10) used as the mobile phase, a sample of 25 µL was inlet into the column and the outlet comes out at 1ml/min flowing capacity. The Peak of analyte was detected at a retention time of 2.037 min. Results: A perfect positive linear relationship between the concentration (10-70μg/mL) and correlation coefficient of 0.999 for Diltiazem was showed by the calibration curve of Diltiazem. This method had a LOD value of 1.276 and a LOQ value of 3.86 respectively. The absolute recovery was 96.52% for Diltiazem tablet. Conclusion: The assay was stability-indicating because the degradation product from stress studies does not affect the detection of diltiazem. Keywords: Diltiazem, RP-HPLC, Methanol, Validation
Authors
- Sudhir Kumar Sahoo (ORCID: https://orcid.org/0000-0002-2222-7957)
- Henerita Dash
Publication Details
- Journal
- Journal of Pharmaceutical Research
- Published
- 2024-12-18
- DOI
- https://doi.org/10.18579/jopcr/v23.4.87
- Primary Topic
- Analytical Methods in Pharmaceuticals
- Type
- article
- Field-Weighted Citation Impact
- 0.00