How Mechanical Properties of Ureteral Stent Material Affect Its Service Life

Ureteral stents are flexible implants commonly made from polyurethane and silicone. Some ureteral stents used in Indonesia experience damage upon removal, even when patients arrive on time for stent removal. This study was conducted to investigate the effect of the mechanical properties of ureteral stent materials on their service life. The tests conducted included tensile testing and material composition analysis using two different Fourier Transform Infrared Spectroscopy (FTIR) devices. In addition, to determine the presence of material mixtures, composition testing was also performed using SEM/EDX. Five samples of ureteral stent products commonly available in Indonesia were selected for this study: two samples with a service life of 2 months, one sample with a 6-month service life, and two samples with a 12-month service life. Analysis of Variance (ANOVA) testing was carried out to confirm correlations between the investigated variables. The results showed that there was no significant effect of the material's mechanical properties, either tensile strength or elastic modulus, on the service life of the ureteral stent. The results showed no statistically significant correlation (p > 0.05) between tensile strength or elastic modulus and service life. Instead, material choice should match intended duration: polyurethane for short-term (< 6 months) and silicone for long-term (> 6 months). Recommended minimum tensile strength is 11.81 MPa for silicone and 9.11 MPa for polyurethane. These results provide practical guidance for designing ureteral stents based on service life rather than mechanical properties.

Authors

Institutions

Publication Details

Journal
Journal of Biomedical Materials Research Part B Applied Biomaterials
Published
2025-11-01
DOI
https://doi.org/10.1002/jbm.b.35692
Primary Topic
Kidney Stones and Urolithiasis Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

How Mechanical Properties of Ureteral Stent Material Affect Its Service Life

Achmad Syaifudin, Lukman Hakim, Katsuhiko SASAKI, Siti Nurkhamidah et al.
Journal of Biomedical Materials Research Part B Applied Biomaterials
Kidney Stones and Urolithiasis Treatments
article

How Mechanical Properties of Ureteral Stent Material Affect Its Service Life

Achmad Syaifudin, Lukman Hakim, Katsuhiko SASAKI, Siti Nurkhamidah, Amaliya Rasyida, Zafarel Zeta Averil, Maulana Y. Izzuddin
article en

Abstract

Ureteral stents are flexible implants commonly made from polyurethane and silicone. Some ureteral stents used in Indonesia experience damage upon removal, even when patients arrive on time for stent removal. This study was conducted to investigate the effect of the mechanical properties of ureteral stent materials on their service life. The tests conducted included tensile testing and material composition analysis using two different Fourier Transform Infrared Spectroscopy (FTIR) devices. In addition, to determine the presence of material mixtures, composition testing was also performed using SEM/EDX. Five samples of ureteral stent products commonly available in Indonesia were selected for this study: two samples with a service life of 2 months, one sample with a 6-month service life, and two samples with a 12-month service life. Analysis of Variance (ANOVA) testing was carried out to confirm correlations between the investigated variables. The results showed that there was no significant effect of the material's mechanical properties, either tensile strength or elastic modulus, on the service life of the ureteral stent. The results showed no statistically significant correlation (p > 0.05) between tensile strength or elastic modulus and service life. Instead, material choice should match intended duration: polyurethane for short-term (< 6 months) and silicone for long-term (> 6 months). Recommended minimum tensile strength is 11.81 MPa for silicone and 9.11 MPa for polyurethane. These results provide practical guidance for designing ureteral stents based on service life rather than mechanical properties.

Journal of Biomedical Materials Research Part B Applied BiomaterialsVol. 113(11)
Hokkaido University of Science (JP), Sepuluh Nopember Institute of Technology (ID), Sapporo University (JP), Universitas Airlangga (ID), Hokkaido University (JP)
Openalex Percentile: Top 51%
Kidney Stones and Urolithiasis Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.