Room return effect of secondary neutrons generated by protons, helium, carbon and oxygen ions for radiotherapy

<i>Objective.</i>Secondary neutrons generated during ion beam radiotherapy undergo scattering from treatment room structures. However, their impact for different primary ion species remains insufficiently characterised. Therefore, this study aims to quantify the room return effect of secondary neutrons in radiotherapy for four different primary ion species, namely protons, helium, carbon and oxygen ions, with energies in the range relevant for radiotherapy.<i>Approach.</i>Ambient dose equivalent,<i>H</i><sup>*</sup>(10), was measured using three types of rem counters to characterise the neutron field generated by mono-energetic beams of increasing energy of the four primary ions. The rem counters were iteratively placed in four positions around a 30 cm × 30 cm × 30 cm RW3 phantom. Experimental data were compared to Monte Carlo (MC) simulation using a detailed room geometry. Next, the simulation was performed without the room to quantify the room return effect.<i>Main results.</i>MC simulations agreed with the experimental data within the uncertainty ranges.<i>H</i><sup>*</sup>(10) decreased with increasing angle relative to the beam direction but increased with higher primary beam energies. Among the ion species studied, oxygen produced the highest values of<i>H</i><sup>*</sup>(10) per primary particle, while protons produced the lowest. The room return effect was found to increase with both, the larger angles from the beam axis and the increasing ion energy, ranging from 17% up to 83% of the total<i>H</i><sup>*</sup>(10).<i>Significance.</i>This study presents the first quantitative assessment of the room return effect for four primary ion species, protons, helium, carbon, and oxygen, for clinically relevant energies. The results demonstrate that the treatment room itself plays a significant role for<i>H</i><sup>*</sup>(10), particularly through contributions from scattered secondary neutrons. Accurate modelling of the room geometry can help improve the reliability of MC simulations and reduce the risk of secondary neutron exposure misestimation during ion beam therapy.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-11-12
DOI
https://doi.org/10.1088/1361-6560/ae1ee6
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Room return effect of secondary neutrons generated by protons, helium, carbon and oxygen ions for radiotherapy

Alberto Stabilini, Christina Stengl, José Vedelago, Federico Alejandro Geser et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

Room return effect of secondary neutrons generated by protons, helium, carbon and oxygen ions for radiotherapy

Alberto Stabilini, Christina Stengl, José Vedelago, Federico Alejandro Geser, J. Mahnke, Phillip J. Haney, Christina Mooshammer, Laura Rozo-Pardo
article en

Abstract

<i>Objective.</i>Secondary neutrons generated during ion beam radiotherapy undergo scattering from treatment room structures. However, their impact for different primary ion species remains insufficiently characterised. Therefore, this study aims to quantify the room return effect of secondary neutrons in radiotherapy for four different primary ion species, namely protons, helium, carbon and oxygen ions, with energies in the range relevant for radiotherapy.<i>Approach.</i>Ambient dose equivalent,<i>H</i><sup>*</sup>(10), was measured using three types of rem counters to characterise the neutron field generated by mono-energetic beams of increasing energy of the four primary ions. The rem counters were iteratively placed in four positions around a 30 cm × 30 cm × 30 cm RW3 phantom. Experimental data were compared to Monte Carlo (MC) simulation using a detailed room geometry. Next, the simulation was performed without the room to quantify the room return effect.<i>Main results.</i>MC simulations agreed with the experimental data within the uncertainty ranges.<i>H</i><sup>*</sup>(10) decreased with increasing angle relative to the beam direction but increased with higher primary beam energies. Among the ion species studied, oxygen produced the highest values of<i>H</i><sup>*</sup>(10) per primary particle, while protons produced the lowest. The room return effect was found to increase with both, the larger angles from the beam axis and the increasing ion energy, ranging from 17% up to 83% of the total<i>H</i><sup>*</sup>(10).<i>Significance.</i>This study presents the first quantitative assessment of the room return effect for four primary ion species, protons, helium, carbon, and oxygen, for clinically relevant energies. The results demonstrate that the treatment room itself plays a significant role for<i>H</i><sup>*</sup>(10), particularly through contributions from scattered secondary neutrons. Accurate modelling of the room geometry can help improve the reliability of MC simulations and reduce the risk of secondary neutron exposure misestimation during ion beam therapy.

Physics in Medicine and BiologyVol. 70(23)
German Cancer Research Center (DE), Heidelberg University (DE), University Hospital Heidelberg (DE), Paul Scherrer Institute (CH)
Openalex Percentile: Top 51%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.