Evaluation of laser Doppler vibrometer’s performance by a multimode laser and changeable reference arm length, equipped with an auto-focus system

In the present study, the theoretical framework underlying interferometric measurement using multimode lasers-forming the basis of laser vibrometry-is comprehensively described. A numerical simulation of the Doppler signal for the Laser Doppler Vibrometer (LDV) is conducted, and the technique for extracting vibration velocity via zero-crossing counting is detailed and subsequently verified through experimental assessment. In addition, two critical optical challenges in this field, namely the determination of the optimal object distances and laser focusing, are addressed. To overcome these issues, a novel set-up is proposed that incorporates systematic adjustments of the reference arm length along with active auto-focusing, facilitated by a laser rangefinder. Moreover, to validate the performance of the proposed device and facilitate comparison with conventional contact accelerometers, several frequency response function (FRF) tests were conducted under different conditions. The test cases included laser measurements when the accelerometer was not mounted on the plate, measurements obtained directly from the accelerometer, laser measurements when the beam was directed onto the accelerometer, and laser measurements when the beam was directed onto the plate. Comparative analysis of the data acquired via the accelerometer and the LDV confirmed the accuracy and reliability of the LDV set-up. Moreover, the measurements revealed that mass loading from the accelerometer caused detectable shifts in the natural frequencies of the 1-mm-thick galvanized plate-shifts that typically require a correction step often treated as modal analysis noise in the literature. This underscores the advantage of LDV's non-contact nature in preserving the structure's true vibrational characteristics of the sample and eliminates the need for such corrective procedures.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-11-13
DOI
https://doi.org/10.1038/s41598-025-23402-3
Primary Topic
Optical measurement and interference techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of laser Doppler vibrometer’s performance by a multimode laser and changeable reference arm length, equipped with an auto-focus system

Fatemeh Rezaei, Farshid Seydipour
Scientific Reports
Optical measurement and interference techniques
article

Evaluation of laser Doppler vibrometer’s performance by a multimode laser and changeable reference arm length, equipped with an auto-focus system

Fatemeh Rezaei, Farshid Seydipour
article en

Abstract

In the present study, the theoretical framework underlying interferometric measurement using multimode lasers-forming the basis of laser vibrometry-is comprehensively described. A numerical simulation of the Doppler signal for the Laser Doppler Vibrometer (LDV) is conducted, and the technique for extracting vibration velocity via zero-crossing counting is detailed and subsequently verified through experimental assessment. In addition, two critical optical challenges in this field, namely the determination of the optimal object distances and laser focusing, are addressed. To overcome these issues, a novel set-up is proposed that incorporates systematic adjustments of the reference arm length along with active auto-focusing, facilitated by a laser rangefinder. Moreover, to validate the performance of the proposed device and facilitate comparison with conventional contact accelerometers, several frequency response function (FRF) tests were conducted under different conditions. The test cases included laser measurements when the accelerometer was not mounted on the plate, measurements obtained directly from the accelerometer, laser measurements when the beam was directed onto the accelerometer, and laser measurements when the beam was directed onto the plate. Comparative analysis of the data acquired via the accelerometer and the LDV confirmed the accuracy and reliability of the LDV set-up. Moreover, the measurements revealed that mass loading from the accelerometer caused detectable shifts in the natural frequencies of the 1-mm-thick galvanized plate-shifts that typically require a correction step often treated as modal analysis noise in the literature. This underscores the advantage of LDV's non-contact nature in preserving the structure's true vibrational characteristics of the sample and eliminates the need for such corrective procedures.

Scientific ReportsVol. 15(1)
K. N. Toosi University of Technology (IR)
Openalex Percentile: Top 65%
Optical measurement and interference techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.