Effects of different drying temperatures and times on the drying efficiency in reprocessing of flexible endoscopes

Background: The drying process is a crucial step in the reprocessing of flexible endoscopes. However, the optimal drying time remains inconclusive, and the impact of temperature on drying efficiency is still unclear. This study aims to provide experimental data on the optimal parameters for endoscope drying in terms of time and temperature. Methods: Ten gastroscopes and ten colonoscopes from a tertiary hospital’s endoscopy center were randomly assigned to different groups based on two temperature conditions (30°C and 50°C) and dried for 30 seconds, 3 minutes, 6 minutes, 9 minutes, and 12 minutes, respectively. Residual liquid was quantitatively assessed using a borescope to determine the optimal drying parameters. Results: A total of 200 observations were made across the temperature and time combinations. The results showed significant differences in the number of residual droplets and the drying pass rates at different times (P < 0.001). A drying time of 12 minutes achieved complete drying of the gastroscopes. Notably, after 30 seconds of drying, a significant difference was observed in the number of residual droplets between the two temperature groups (P < 0.05), with fewer droplets present at 50°C. Conclusion: Drying time significantly affects the drying efficiency of endoscopes, with 12 minutes being the optimal duration. Maintaining a temperature of 50°C can shorten drying time and improve efficiency.

Authors

Institutions

Publication Details

Journal
Saudi Journal of Gastroenterology
Published
2025-11-12
DOI
https://doi.org/10.4103/sjg.sjg_45_25
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of different drying temperatures and times on the drying efficiency in reprocessing of flexible endoscopes

Xun-Zhang, Jiuhong Ma, Huang Xi, Chen Liu-Ji et al.
Saudi Journal of Gastroenterology
Medical Device Sterilization and Disinfection
article

Effects of different drying temperatures and times on the drying efficiency in reprocessing of flexible endoscopes

Xun-Zhang, Jiuhong Ma, Huang Xi, Chen Liu-Ji, Zhen Jing, Li Xian-Huang, Liu Hai-Xia, Zhang Yan-Xia, Huang Zi-Yi
article en

Abstract

Background: The drying process is a crucial step in the reprocessing of flexible endoscopes. However, the optimal drying time remains inconclusive, and the impact of temperature on drying efficiency is still unclear. This study aims to provide experimental data on the optimal parameters for endoscope drying in terms of time and temperature. Methods: Ten gastroscopes and ten colonoscopes from a tertiary hospital’s endoscopy center were randomly assigned to different groups based on two temperature conditions (30°C and 50°C) and dried for 30 seconds, 3 minutes, 6 minutes, 9 minutes, and 12 minutes, respectively. Residual liquid was quantitatively assessed using a borescope to determine the optimal drying parameters. Results: A total of 200 observations were made across the temperature and time combinations. The results showed significant differences in the number of residual droplets and the drying pass rates at different times (P < 0.001). A drying time of 12 minutes achieved complete drying of the gastroscopes. Notably, after 30 seconds of drying, a significant difference was observed in the number of residual droplets between the two temperature groups (P < 0.05), with fewer droplets present at 50°C. Conclusion: Drying time significantly affects the drying efficiency of endoscopes, with 12 minutes being the optimal duration. Maintaining a temperature of 50°C can shorten drying time and improve efficiency.

Saudi Journal of GastroenterologyVol. 32(1)
Nanchang University (CN), Jiangxi University of Technology (CN), First Affiliated Hospital of Jiangxi Medical College (CN), Jiangxi Agricultural University (CN)
Openalex Percentile: Top 68%
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.