Pleural malpositioning of a haemodialysis catheter causing haemothorax despite the use of ultrasound guided placement and chest radiograph confirmation.

We report an unusual case of haemodialysis catheter malposition into the pleural space causing haemothorax, despite ultrasound-guided insertion and X-ray confirmation. A patient with chronic kidney disease presented in respiratory failure due to pulmonary oedema and was scheduled for urgent haemodialysis. A haemodialysis catheter was inserted under ultrasound guidance, with apparent blood backflow, and postinsertion chest radiograph suggested proper placement. However, dialysis was interrupted multiple times due to persistently inadequate extracorporeal blood flow and visibly diluted blood in the circuit. Blood samples from the catheter revealed a significant drop in haemoglobin levels compared with baseline. Further imaging revealed catheter misplacement into the pleural cavity with haemothorax. This case underscores that even with standard placement verification techniques, rare malposition can occur. Clinicians should maintain a high index of suspicion and pursue further evaluation when unexplained flow or clinical abnormalities arise during dialysis.

Authors

Institutions

Publication Details

Journal
BMJ Case Reports
Published
2025-11-01
DOI
https://doi.org/10.1136/bcr-2025-267403
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pleural malpositioning of a haemodialysis catheter causing haemothorax despite the use of ultrasound guided placement and chest radiograph confirmation.

Sireesha Koneru, Venkat Raman Kola, Ravi Shankar Vattikuti, Madhava Reddy Devireddy
BMJ Case Reports
Central Venous Catheters and Hemodialysis
article

Pleural malpositioning of a haemodialysis catheter causing haemothorax despite the use of ultrasound guided placement and chest radiograph confirmation.

Sireesha Koneru, Venkat Raman Kola, Ravi Shankar Vattikuti, Madhava Reddy Devireddy
article en

Abstract

We report an unusual case of haemodialysis catheter malposition into the pleural space causing haemothorax, despite ultrasound-guided insertion and X-ray confirmation. A patient with chronic kidney disease presented in respiratory failure due to pulmonary oedema and was scheduled for urgent haemodialysis. A haemodialysis catheter was inserted under ultrasound guidance, with apparent blood backflow, and postinsertion chest radiograph suggested proper placement. However, dialysis was interrupted multiple times due to persistently inadequate extracorporeal blood flow and visibly diluted blood in the circuit. Blood samples from the catheter revealed a significant drop in haemoglobin levels compared with baseline. Further imaging revealed catheter misplacement into the pleural cavity with haemothorax. This case underscores that even with standard placement verification techniques, rare malposition can occur. Clinicians should maintain a high index of suspicion and pursue further evaluation when unexplained flow or clinical abnormalities arise during dialysis.

BMJ Case ReportsVol. 18(11)
Yashoda Hospital (IN)
Openalex Percentile: Top 49%
Central Venous Catheters and Hemodialysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.