A figure of merit for transcranial ultrasound transmission based on coupled elastic-thermal analysis

<i>Objective.</i>To evaluate the efficiency of ultrasound transmission and its associated thermal effects on the skull, introducing a figure of merit (FOM) to assess the efficacy and safety of transcranial focused ultrasound (tFUS) at varying incidence angles and frequencies.<i>Approach</i>. A coupled elastic-thermal analysis was performed using a two-dimensional (2D) parallel solid plate skull model, validated through three-dimensional simulations. High-resolution micro-computed tomography (microCT) data from an<i>ex vivo</i>human skull sample was incorporated for realistic modeling. Simulations were conducted at frequencies of 0.8 MHz, 1 MHz, and 1.2 MHz to explore the impact of skull thickness, incidence angle, and absorption properties on ultrasound transmission and heating. The FOM was calculated for each scenario to provide an integrated measure of transmission efficiency and thermal safety. Experimental studies using the same<i>ex vivo</i>human skull validated the microCT-based simulation model.<i>Main results</i>. The FOM analysis demonstrated that at 0.8 MHz for an 8 mm thick skull, the FOM dropped by half for incidence angles above 15°, indicating reduced ultrasound transmission and heightened temperature rise at oblique angles. These findings suggest that dual-mode conversion at higher angles may not be suitable for tFUS applications, supporting the clinical preference for near-normal incidence. The validated 2D microCT model provides a reliable framework for studying the interplay between transmission and heating effects.<i>Significance</i>. This study provides an integrated framework for optimizing tFUS parameters by balancing transmission efficiency and thermal safety. The introduced FOM offers a novel metric to guide therapeutic applications, contributing to safer and more effective transcranial ultrasound treatments.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-11-12
DOI
https://doi.org/10.1088/1361-6560/ae1ee3
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A figure of merit for transcranial ultrasound transmission based on coupled elastic-thermal analysis

Costas Arvanitis, M. Sait Kilinc, F. Levent Degertekin
Physics in Medicine and Biology
Ultrasound and Hyperthermia Applications
article

A figure of merit for transcranial ultrasound transmission based on coupled elastic-thermal analysis

Costas Arvanitis, M. Sait Kilinc, F. Levent Degertekin
article en

Abstract

<i>Objective.</i>To evaluate the efficiency of ultrasound transmission and its associated thermal effects on the skull, introducing a figure of merit (FOM) to assess the efficacy and safety of transcranial focused ultrasound (tFUS) at varying incidence angles and frequencies.<i>Approach</i>. A coupled elastic-thermal analysis was performed using a two-dimensional (2D) parallel solid plate skull model, validated through three-dimensional simulations. High-resolution micro-computed tomography (microCT) data from an<i>ex vivo</i>human skull sample was incorporated for realistic modeling. Simulations were conducted at frequencies of 0.8 MHz, 1 MHz, and 1.2 MHz to explore the impact of skull thickness, incidence angle, and absorption properties on ultrasound transmission and heating. The FOM was calculated for each scenario to provide an integrated measure of transmission efficiency and thermal safety. Experimental studies using the same<i>ex vivo</i>human skull validated the microCT-based simulation model.<i>Main results</i>. The FOM analysis demonstrated that at 0.8 MHz for an 8 mm thick skull, the FOM dropped by half for incidence angles above 15°, indicating reduced ultrasound transmission and heightened temperature rise at oblique angles. These findings suggest that dual-mode conversion at higher angles may not be suitable for tFUS applications, supporting the clinical preference for near-normal incidence. The validated 2D microCT model provides a reliable framework for studying the interplay between transmission and heating effects.<i>Significance</i>. This study provides an integrated framework for optimizing tFUS parameters by balancing transmission efficiency and thermal safety. The introduced FOM offers a novel metric to guide therapeutic applications, contributing to safer and more effective transcranial ultrasound treatments.

Physics in Medicine and BiologyVol. 70(23)
Georgia Institute of Technology (US), The Wallace H. Coulter Department of Biomedical Engineering (US)
Openalex Percentile: Top 67%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.