Development of a Calibration Transfer Methodology and Experimental Setup for Urine Headspace Analysis

Electronic noses (E-Noses) equipped with metal-oxide semiconductor (MOS) sensors are promising tools for non-invasive medical diagnostics. Their adoption in clinical practice, however, is limited—among others—by sensor variability across devices, which makes individual calibration necessary. This study presents an approach for the development of a calibration transfer (CT) methodology for urine headspace analysis, involving the design and realization of a dedicated experimental setup and protocol. Partial least squares-discriminant analysis (PLS-DA) models were trained on human urine samples enriched with selected biomarkers to simulate pathological states. Models from a reference (“master”) device were transferred to other (“slave”) units in multiple master–slave configurations using Direct Standardization (DS). To overcome the variability of human urine, synthetic urine recipes were formulated to mimic sensor responses and serve as reproducible transfer samples. Several strategies for selecting transfer samples were evaluated, including the Kennard–Stone algorithm, a DBSCAN-based approach, and random selection. Without CT, classification accuracy on slave devices decreased markedly (37–55%) compared to the master’s performance (79%), whereas applying DS with synthetic standards restored accuracy to 75–80%. These results demonstrate that combining reproducible synthetic standards with DS enables effective model transfer across E-Noses, reducing calibration requirements and supporting their broader applicability in medical diagnostics.

Authors

Institutions

Publication Details

Journal
Chemosensors
Published
2025-11-12
DOI
https://doi.org/10.3390/chemosensors13110395
Citations
3
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
2.02
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a Calibration Transfer Methodology and Experimental Setup for Urine Headspace Analysis

Laura Capelli, Raffaele L. Dellacá, Beatrice Julia Lotesoriere, Fabio Grizzi et al.
3 citations
Chemosensors
Advanced Chemical Sensor Technologies
2.02
article

Development of a Calibration Transfer Methodology and Experimental Setup for Urine Headspace Analysis

Laura Capelli, Raffaele L. Dellacá, Beatrice Julia Lotesoriere, Fabio Grizzi, Gianluigi Taverna, Stefano Robbiani, Marco Cassinerio, Emanuele Zanni
article en
3 citations

Abstract

Electronic noses (E-Noses) equipped with metal-oxide semiconductor (MOS) sensors are promising tools for non-invasive medical diagnostics. Their adoption in clinical practice, however, is limited—among others—by sensor variability across devices, which makes individual calibration necessary. This study presents an approach for the development of a calibration transfer (CT) methodology for urine headspace analysis, involving the design and realization of a dedicated experimental setup and protocol. Partial least squares-discriminant analysis (PLS-DA) models were trained on human urine samples enriched with selected biomarkers to simulate pathological states. Models from a reference (“master”) device were transferred to other (“slave”) units in multiple master–slave configurations using Direct Standardization (DS). To overcome the variability of human urine, synthetic urine recipes were formulated to mimic sensor responses and serve as reproducible transfer samples. Several strategies for selecting transfer samples were evaluated, including the Kennard–Stone algorithm, a DBSCAN-based approach, and random selection. Without CT, classification accuracy on slave devices decreased markedly (37–55%) compared to the master’s performance (79%), whereas applying DS with synthetic standards restored accuracy to 75–80%. These results demonstrate that combining reproducible synthetic standards with DS enables effective model transfer across E-Noses, reducing calibration requirements and supporting their broader applicability in medical diagnostics.

ChemosensorsVol. 13(11)
Humanitas Mater Domini (IT), IRCCS Humanitas Research Hospital (IT), Politecnico di Milano (IT)
Openalex Percentile: Top 12%
Advanced Chemical Sensor Technologies
2.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.