Towards Realistic 3D-Printed Phantoms for Aneurysm Clipping: Mechanical Characterisation of Basilar Arteries

Cerebral aneurysm clipping remains a key surgical approach despite advancements in endovascular techniques. However, training for this procedure is complex due to the variable and fragile nature of aneurysmal tissues. This study evaluates the mechanical behaviour of human basilar arteries during clipping and compares them to 3D-printed models used for neurosurgical training. Mechanical tests were performed on ten cadaveric basilar arteries, distinguishing between healthy and plaque-affected segments. Plaque-affected regions required significantly higher clipping force (1.73 ± 0.22 N) compared to healthy segments (0.45 ± 0.19 N), confirming that atherosclerosis markedly increases arterial stiffness. Six 3D-printed phantom materials were evaluated; none accurately replicated the biomechanical response of real arteries. The Flex Anatomical material showed the highest stiffness (44.51 ± 0.98 N), while Silicone 40A was the most compliant (1.05 ± 0.12 N), yet both deviated substantially from biological performance. These findings underscore the current limitations of anatomical models that lack realistic biomechanical properties.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2025-11-12
DOI
https://doi.org/10.3390/bioengineering12111239
Primary Topic
Anatomy and Medical Technology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Towards Realistic 3D-Printed Phantoms for Aneurysm Clipping: Mechanical Characterisation of Basilar Arteries

Lukáš Čapek, Jiří Šafka, Petra Hamrikova, Pavel Buchvald et al.
Bioengineering
Anatomy and Medical Technology
article

Towards Realistic 3D-Printed Phantoms for Aneurysm Clipping: Mechanical Characterisation of Basilar Arteries

Lukáš Čapek, Jiří Šafka, Petra Hamrikova, Pavel Buchvald, Jiří Vitvar
article en

Abstract

Cerebral aneurysm clipping remains a key surgical approach despite advancements in endovascular techniques. However, training for this procedure is complex due to the variable and fragile nature of aneurysmal tissues. This study evaluates the mechanical behaviour of human basilar arteries during clipping and compares them to 3D-printed models used for neurosurgical training. Mechanical tests were performed on ten cadaveric basilar arteries, distinguishing between healthy and plaque-affected segments. Plaque-affected regions required significantly higher clipping force (1.73 ± 0.22 N) compared to healthy segments (0.45 ± 0.19 N), confirming that atherosclerosis markedly increases arterial stiffness. Six 3D-printed phantom materials were evaluated; none accurately replicated the biomechanical response of real arteries. The Flex Anatomical material showed the highest stiffness (44.51 ± 0.98 N), while Silicone 40A was the most compliant (1.05 ± 0.12 N), yet both deviated substantially from biological performance. These findings underscore the current limitations of anatomical models that lack realistic biomechanical properties.

BioengineeringVol. 12(11)
Technical University of Liberec (CZ), Krajská Nemocnice Liberec (CZ), Nanomaterials Research (United States) (US), University Hospital Hradec Králové (CZ), University of Hradec Králové (CZ)
Openalex Percentile: Top 67%
Anatomy and Medical Technology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.