Newly developed plastic stent delivery system for endoscopic ultrasonography-guided gallbladder drainage: Experiments on gallbladder models
Endoscopic ultrasonography-guided gallbladder drainage (EUS-GBD) using the current plastic stent (PS) system requires multiple steps for stent deployment and carries the risk of bile leakage from the puncture site. In this study, we developed a single-step PS system that includes a 3F electrocautery metal tip on the inner sheath mounted with a 7F/4-cm double-pigtail PS and experimentally evaluated it on a gallbladder model mimicking EUS-GBD. We assessed procedure time, intra-gallbladder pressure-drop ratios, and bile leakage using a newly developed single-step PS system, comparing it with the current PS system during drainage attempts on a gallbladder model. Mean duration times for the three attempts were 2 minutes 59 seconds and 27 seconds for the current PS system and the newly developed PS system groups, respectively ( <i>P</i> < 0.001). Mean intra-gallbladder pressure-drop ratios were 86.7% in the current PS system group and 7.6% in the newly developed PS system group ( <i>P</i> < 0.001). The newly developed PS system prevented bile leakage from the puncture site. The newly developed PS system saved significant time, maintained intra-gallbladder pressure, and prevented bile leakage during the procedure when compared with the current PS system.
Authors
- Tesshin Ban (ORCID: https://orcid.org/0000-0001-8823-1476)
- Takashi Joh
- Yoshimasa Kubota (ORCID: https://orcid.org/0000-0002-6475-8035)
Institutions
- Kumamoto City Hospital (JP)
Publication Details
- Journal
- Endoscopy International Open
- Published
- 2025-11-11
- DOI
- https://doi.org/10.1055/a-2734-0383
- Primary Topic
- Gallbladder and Bile Duct Disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.00