Eye lens dose and shielding efficacy of radiation-protective eyewear in C-arm–guided cervical and lumbar epidural blocks: a phantom study

Epidural blocks are commonly performed interventional pain procedures under C-arm fluoroscopic guidance. However, radiation exposure to the ocular lens remains a significant concern due to its high radiosensitivity and the associated risk of radiation-induced cataract formation. Therefore, this study aims to quantify ocular lens dose during cervical and lumbar epidural blocks (CEBs and LEBs), and to evaluate the shielding efficacy of 0.5 mmPb lead-free radiation-protective eyewear (R-PEW), using a quantitative, phantom-based experimental design that simulates realistic clinical scenarios based on exposure conditions derived from 150 real patients. An adult anthropomorphic phantom was used, and ocular lens doses were recorded with optically stimulated luminescence dosimeters under six experimental conditions defined based on procedure type, patient position, and eyewear status. Without shielding, the maximum lens dose was 1.90 mSv during prone CEBs, while the greatest reduction with R-PEW was 83.3% during prone LEBs. Dose asymmetry between the right and left eyes corresponded to differences in source-to-eye distance. All measured doses were substantially below the cataract threshold (0.5 Gy) and the International Commission on Radiological Protection annual public equivalent dose limit for the lens of the eye (15 mSv). Nonetheless, R-PEW consistently reduced exposure under all conditions, offering additional protection in accordance with the optimisation principle. These findings suggest that, although routine patient shielding is not warranted given the low absolute doses, R-PEW represents a simple and effective adjunct to further minimise ocular exposure during C-arm-guided epidural blocks.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2025-11-11
DOI
https://doi.org/10.1088/1361-6498/ae1e05
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Eye lens dose and shielding efficacy of radiation-protective eyewear in C-arm–guided cervical and lumbar epidural blocks: a phantom study

D. Y. Rhee, 지원 김, Jinseok Yeo, Hee-June Kim et al.
Journal of Radiological Protection
Radiation Dose and Imaging
article

Eye lens dose and shielding efficacy of radiation-protective eyewear in C-arm–guided cervical and lumbar epidural blocks: a phantom study

D. Y. Rhee, 지원 김, Jinseok Yeo, Hee-June Kim, 일환 배
article en

Abstract

Epidural blocks are commonly performed interventional pain procedures under C-arm fluoroscopic guidance. However, radiation exposure to the ocular lens remains a significant concern due to its high radiosensitivity and the associated risk of radiation-induced cataract formation. Therefore, this study aims to quantify ocular lens dose during cervical and lumbar epidural blocks (CEBs and LEBs), and to evaluate the shielding efficacy of 0.5 mmPb lead-free radiation-protective eyewear (R-PEW), using a quantitative, phantom-based experimental design that simulates realistic clinical scenarios based on exposure conditions derived from 150 real patients. An adult anthropomorphic phantom was used, and ocular lens doses were recorded with optically stimulated luminescence dosimeters under six experimental conditions defined based on procedure type, patient position, and eyewear status. Without shielding, the maximum lens dose was 1.90 mSv during prone CEBs, while the greatest reduction with R-PEW was 83.3% during prone LEBs. Dose asymmetry between the right and left eyes corresponded to differences in source-to-eye distance. All measured doses were substantially below the cataract threshold (0.5 Gy) and the International Commission on Radiological Protection annual public equivalent dose limit for the lens of the eye (15 mSv). Nonetheless, R-PEW consistently reduced exposure under all conditions, offering additional protection in accordance with the optimisation principle. These findings suggest that, although routine patient shielding is not warranted given the low absolute doses, R-PEW represents a simple and effective adjunct to further minimise ocular exposure during C-arm-guided epidural blocks.

Journal of Radiological ProtectionVol. 45(4)
Kyungpook National University Hospital (KR), Daegu Catholic University (KR), Kangwon National University Hospital (KR)
Openalex Percentile: Top 51%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.