Initial experience with LVIS <sup>TM</sup> EVO <sup>TM</sup> for distal flow diverting effects

BackgroundFlow diversion for intracranial aneurysms is indicated for wide-necked aneurysms in non-perforator-rich segments. However, devices are challenging to deploy in distal/tortuous vasculature. LVIS<sup>TM</sup> EVO<sup>TM</sup> is a braided intermediate-density stent offering greater flexibility and navigability than flow diverters with higher metal-to-surface ratios. We investigated its off-label use as an independent flow diverter for small, complex intracranial aneurysms.MethodsWe retrospectively reviewed patients at a single center who underwent independent stenting with LVIS<sup>TM</sup> EVO<sup>TM</sup> from December 2023 to September 2024. Demographics, procedural details, and angiographic outcomes were analyzed. Follow-up computed tomography angiography was performed from May 2024 to October 2025. Descriptive statistics were used; non-parametric tests compared aneurysm and parent vessel diameters.ResultsNine patients (median age 59; 89% female) with off-label intracranial aneurysms (median diameter 5 mm) were treated using LVIS<sup>TM</sup> EVO<sup>TM</sup> (median diameter 3 mm, length 18 mm). Deployment was technically successful in 100% of cases with no peri-procedural complications. One patient experienced delayed hemorrhage with full neurological recovery. At follow-up (available for 6 patients), 83.3% showed complete aneurysm occlusion (Raymond-Roy Occlusion Classification [RROC] class I). All patients demonstrated 100% stent patency and no new neurological deficits.ConclusionLVIS<sup>TM</sup> EVO<sup>TM</sup> demonstrated technical feasibility and acceptable initial safety as an off-label flow-diverting device for small (as low as 2.5 mm), distal aneurysms in this limited single-center experience. Its favorable deliverability, safety profile, and occlusion rates suggest it as a possible alternative to currently marketed flow diverters in anatomically challenging patients. These findings warrant validation in larger, multicenter studies with extended follow-up.

Authors

Institutions

Publication Details

Journal
Interventional Neuroradiology
Published
2025-11-11
DOI
https://doi.org/10.1177/15910199251394570
Primary Topic
Intracranial Aneurysms: Treatment and Complications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Initial experience with LVIS TM EVO TM for distal flow diverting effects

Y. Jonathan Zhang, Ferdinand Hui, Aditi Suri, Yang Qiao et al.
Interventional Neuroradiology
Intracranial Aneurysms: Treatment and Complications
article

Initial experience with LVIS TM EVO TM for distal flow diverting effects

Y. Jonathan Zhang, Ferdinand Hui, Aditi Suri, Yang Qiao, Trevor Torigoe, S Tsappidi
article en

Abstract

BackgroundFlow diversion for intracranial aneurysms is indicated for wide-necked aneurysms in non-perforator-rich segments. However, devices are challenging to deploy in distal/tortuous vasculature. LVIS<sup>TM</sup> EVO<sup>TM</sup> is a braided intermediate-density stent offering greater flexibility and navigability than flow diverters with higher metal-to-surface ratios. We investigated its off-label use as an independent flow diverter for small, complex intracranial aneurysms.MethodsWe retrospectively reviewed patients at a single center who underwent independent stenting with LVIS<sup>TM</sup> EVO<sup>TM</sup> from December 2023 to September 2024. Demographics, procedural details, and angiographic outcomes were analyzed. Follow-up computed tomography angiography was performed from May 2024 to October 2025. Descriptive statistics were used; non-parametric tests compared aneurysm and parent vessel diameters.ResultsNine patients (median age 59; 89% female) with off-label intracranial aneurysms (median diameter 5 mm) were treated using LVIS<sup>TM</sup> EVO<sup>TM</sup> (median diameter 3 mm, length 18 mm). Deployment was technically successful in 100% of cases with no peri-procedural complications. One patient experienced delayed hemorrhage with full neurological recovery. At follow-up (available for 6 patients), 83.3% showed complete aneurysm occlusion (Raymond-Roy Occlusion Classification [RROC] class I). All patients demonstrated 100% stent patency and no new neurological deficits.ConclusionLVIS<sup>TM</sup> EVO<sup>TM</sup> demonstrated technical feasibility and acceptable initial safety as an off-label flow-diverting device for small (as low as 2.5 mm), distal aneurysms in this limited single-center experience. Its favorable deliverability, safety profile, and occlusion rates suggest it as a possible alternative to currently marketed flow diverters in anatomically challenging patients. These findings warrant validation in larger, multicenter studies with extended follow-up.

Interventional Neuroradiology
University of Hawaiʻi at Mānoa (US), Queen's Medical Center (US)
Openalex Percentile: Top 53%
Intracranial Aneurysms: Treatment and Complications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines