Out-of-field dosimetry in VMAT for medulloblastoma using OSLDs
Volumetric modulated arc therapy for pediatric medulloblastoma improves target precision but increases low-dose exposure to healthy tissues, elevating secondary cancer risks. This study evaluates treatment planning system (TPS) accuracy for out-of-field dose (OOFD) predictions by performing measurements with optically stimulated luminescence dosemeters (OSLDs) in an anthropomorphic phantom. TPS predictions were accurate near the field (<3% at 5.5 cm) but significantly underestimated doses at larger distances, with discrepancies ranging from 16% (thyroid) to 64% (heart). These findings highlight TPS limitations in modeling peripheral leakage and scatter and underscore the critical role of OSLD-based monitoring for pediatric patients. Accurate OOFD verification is essential for risk assessment and long-term safety. Future work should focus on refining TPS algorithms and integrating direct dosimetry into clinical practice for this vulnerable population.
Authors
- A. Elanique (ORCID: https://orcid.org/0009-0002-6530-9108)
- Mustapha Krim (ORCID: https://orcid.org/0000-0001-9938-2843)
- I. Mouhti (ORCID: https://orcid.org/0000-0003-2924-8192)
- A. Saadi
- M.R. Mesradi (ORCID: https://orcid.org/0000-0002-7961-3656)
- Yassine Aguezzoumen
- Oualid Warimezgane
- Mohamed Rida Abbari
Institutions
- Université Ibn Zohr (MA)
- Université Hassan 1er (MA)
Publication Details
- Journal
- Radiation Protection Dosimetry
- Published
- 2025-11-11
- DOI
- https://doi.org/10.1093/rpd/ncaf130
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00