Differences in Impulse Oscillometry Parameters Measured Using Different Devices
Significant differences in IOS reactance values were found between the MasterScreen-IOS and MostGraph-02. Device-specific reference equations may contribute to differences in percent predicted values. Therefore, IOS measurements from these devices should be interpreted cautiously and not considered interchangeable for longitudinal assessments.