Transient R-wave amplitude attenuation owing to air entrapment during extravascular implantable cardioverter-defibrillator implantation: Delayed reassessment as a potential bail-out strategy
Key Teaching Points Air entrapment in the substernal space may transiently reduce R-wave sensing during EV-ICD implantation. In this setting, R-wave sensing and defibrillation testing failure may resolve spontaneously as the entrapped air dissipates over time. In borderline cases, delayed reassessment, rather than immediate system revision, may be a feasible and appropriate strategy when air entrapment is involved.