Significance of positional correction of planned couch angles with the ExacTrac system

This study evaluated the benefits of the ExacTrac system for non-coplanar beams in brain stereotactic radiotherapy (SRT) by comparing pre- and post-X-ray verification (XV) positioning data. Data from 62 patients undergoing brain SRT were analyzed, focusing on 151 non-coplanar beams. Patient positioning involved initial infrared marker alignment, followed by ExacTrac kV x-ray acquisition and 6D robotic couch adjustments. Patient positioning was corrected within a tolerance of 0.5 mm and 0.5° for translation and rotation, respectively. Pre and post-XV positioning data were compared at each planned couch angle. Results showed that 54% of non-coplanar beams had positioning errors exceeding the specified tolerances on the pre-XV. Significant differences were observed in translational and rotational corrections between pre- and post-XV positioning at each planned couch angle. These results demonstrate that XV at any each planned couch angle using the ExacTrac system significantly improves patient positioning accuracy in brain SRT.

Authors

Institutions

Publication Details

Journal
Reports of Practical Oncology & Radiotherapy
Published
2025-10-03
DOI
https://doi.org/10.5603/rpor.106488
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Significance of positional correction of planned couch angles with the ExacTrac system

Soichiro Ishihara, Masahiro Kenjo, Shuichi Ozawa, Minoru Nakao et al.
Reports of Practical Oncology & Radiotherapy
Advanced Radiotherapy Techniques
article

Significance of positional correction of planned couch angles with the ExacTrac system

Soichiro Ishihara, Masahiro Kenjo, Shuichi Ozawa, Minoru Nakao, Masayuki Kagemoto, Hideharu Miura
article en

Abstract

This study evaluated the benefits of the ExacTrac system for non-coplanar beams in brain stereotactic radiotherapy (SRT) by comparing pre- and post-X-ray verification (XV) positioning data. Data from 62 patients undergoing brain SRT were analyzed, focusing on 151 non-coplanar beams. Patient positioning involved initial infrared marker alignment, followed by ExacTrac kV x-ray acquisition and 6D robotic couch adjustments. Patient positioning was corrected within a tolerance of 0.5 mm and 0.5° for translation and rotation, respectively. Pre and post-XV positioning data were compared at each planned couch angle. Results showed that 54% of non-coplanar beams had positioning errors exceeding the specified tolerances on the pre-XV. Significant differences were observed in translational and rotational corrections between pre- and post-XV positioning at each planned couch angle. These results demonstrate that XV at any each planned couch angle using the ExacTrac system significantly improves patient positioning accuracy in brain SRT.

Reports of Practical Oncology & RadiotherapyVol. 30(4)
Hiroshima University (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 57%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.