Evaluation of Wideband Pulse Sequences for Suppressing Image Artifacts in Children with Cardiac Implantable Electronic Devices

Purpose To determine whether wideband late gadolinium enhancement (LGE) and wideband perfusion pulse sequences suppress image artifacts in children in the presence of a cardiac implantable electronic device (CIED). Materials and Methods In the first experiment of this prospective study, unenhanced cardiac MRI was performed at 1.5 T in 18 healthy pediatric participants (median age, 12 years [range, 10-15 years]; 12 [67%] female) with and without a CIED generator taped in the subclavicular and abdominal positions. LGE and perfusion scans were performed with standard and wideband sequences in each state. Three clinical raters adjudicated the presence or absence of clinically relevant image artifacts. Image artifacts were also quantified. In the second experiment, wideband perfusion and LGE MRI scans with gadolinium-based contrast material administration were performed on four pediatric participants (median age, 15 years [range, 15-19 years]; three [75%] female) with CIEDs in situ. Image quality was graded on a five-point Likert scale by clinical raters (3 = clinically acceptable). Myocardial blood flow was also quantified. Results In the participant experiments, standard LGE produced image artifacts in 61.81% (178 of 288) and 37.15% (107 of 288) of myocardial segments for subclavicular and abdominal IPGs, respectively, and were significantly different (<i>P</i> < .001) from wideband LGE (14.68% [42 of 288] and 0%, respectively) for both positions. Quantitative analysis results were consistent with expert adjudication. Wideband cardiac MRI in four children with CIEDs achieved high image quality (median score ≥ 4.0) and mean resting myocardial blood flow value of 1.06 mL/min/g (<i>n</i> = 3). Conclusion Compared with standard LGE and perfusion MRI, wideband LGE and perfusion MRI produced significantly fewer image artifacts induced by CIEDs in pediatric participants. <b>Keywords:</b> Wideband, Pediatric, Late Gadolinium Enhancement, Perfusion, Pacemaker, Defibrillator <i>Supplemental material is available for this article.</i> © RSNA, 2025.

Authors

Institutions

Publication Details

Journal
Radiology Cardiothoracic Imaging
Published
2025-11-06
DOI
https://doi.org/10.1148/ryct.250310
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Wideband Pulse Sequences for Suppressing Image Artifacts in Children with Cardiac Implantable Electronic Devices

Andrada Popescu, Gregory Webster, Oluyemi Aboyewa, Christina Laternser et al.
Radiology Cardiothoracic Imaging
Advanced MRI Techniques and Applications
article

Evaluation of Wideband Pulse Sequences for Suppressing Image Artifacts in Children with Cardiac Implantable Electronic Devices

Andrada Popescu, Gregory Webster, Oluyemi Aboyewa, Christina Laternser, Dhaivat Shah, Laleh Golestanirad, KyungPyo Hong, Joseph Camarda, Daniel Kim, Simon Lee
article en

Abstract

Purpose To determine whether wideband late gadolinium enhancement (LGE) and wideband perfusion pulse sequences suppress image artifacts in children in the presence of a cardiac implantable electronic device (CIED). Materials and Methods In the first experiment of this prospective study, unenhanced cardiac MRI was performed at 1.5 T in 18 healthy pediatric participants (median age, 12 years [range, 10-15 years]; 12 [67%] female) with and without a CIED generator taped in the subclavicular and abdominal positions. LGE and perfusion scans were performed with standard and wideband sequences in each state. Three clinical raters adjudicated the presence or absence of clinically relevant image artifacts. Image artifacts were also quantified. In the second experiment, wideband perfusion and LGE MRI scans with gadolinium-based contrast material administration were performed on four pediatric participants (median age, 15 years [range, 15-19 years]; three [75%] female) with CIEDs in situ. Image quality was graded on a five-point Likert scale by clinical raters (3 = clinically acceptable). Myocardial blood flow was also quantified. Results In the participant experiments, standard LGE produced image artifacts in 61.81% (178 of 288) and 37.15% (107 of 288) of myocardial segments for subclavicular and abdominal IPGs, respectively, and were significantly different (<i>P</i> < .001) from wideband LGE (14.68% [42 of 288] and 0%, respectively) for both positions. Quantitative analysis results were consistent with expert adjudication. Wideband cardiac MRI in four children with CIEDs achieved high image quality (median score ≥ 4.0) and mean resting myocardial blood flow value of 1.06 mL/min/g (<i>n</i> = 3). Conclusion Compared with standard LGE and perfusion MRI, wideband LGE and perfusion MRI produced significantly fewer image artifacts induced by CIEDs in pediatric participants. <b>Keywords:</b> Wideband, Pediatric, Late Gadolinium Enhancement, Perfusion, Pacemaker, Defibrillator <i>Supplemental material is available for this article.</i> © RSNA, 2025.

Radiology Cardiothoracic ImagingVol. 7(6)
Northwestern University (US), Lurie Children's Hospital (US), Lurie Children's Hospital (US)
Openalex Percentile: Top 52%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.