Comparative evaluation of diagnostic and analytical performance of DIRUI FUS-200 and MUS-3600 fully automated urine analyzers

The aim was to assess and compare the analytical performance of the FUS-200 and its new upgrade, the MUS-3600 urine autoanalyzers, and to compare performance with manual microscopy. First morning, void urine samples were randomly collected, and suitable samples were analyzed on both autoanalyzers. In addition, concurrent manual microscopic examinations were performed for all suitable urine samples. Carry-over, linearity and imprecision analysis were performed to assess analytical and diagnostic performance of both urine autoanalyzers according to Clinical and Laboratory Standards Institute (CLSI) EP15-A2, and the study was conducted in accordance with CLSI GP16-A3 Urinalysis; Approved Guideline. A total of 518 samples were collected, and of these, 494 (95.4%) were suitable for analysis and were included in the study. Sensitivity and positive likelihood ratio (LR+) values for WBC and squamous epithelium (SqEC) counts for both autoanalyzers were >90% and >10%, respectively. Specificity and LR+ values obtained from MUS-3600 were better compared to the FUS-200 (respectively 92.8 vs. 84.7; 5.5 vs. 10.1). One group agreement between the MUS-3600 and manual microscopy was >90%. Both analyzers displayed comparable performance for RBC and WBC counts, which were moderately correlated with each other. These results show the diagnostic performance of the MUS-3600 and FUS-200 was satisfactory for urine sediment analysis and was compatible with manual microscopy findings. However, random urine samples are sub-optimal for evaluating diagnostic performance, particularly for bacterial, cast, crystal and yeast analysis. Thus, we recommend using more appropriate urine samples for this comparison, such as from patients with suspected urinary tract infections.

Authors

Institutions

Publication Details

Journal
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2025-10-03
DOI
https://doi.org/10.1080/00365513.2025.2582800
Primary Topic
Bacterial Identification and Susceptibility Testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative evaluation of diagnostic and analytical performance of DIRUI FUS-200 and MUS-3600 fully automated urine analyzers

Fatih Hunc, Berrin Öztaş, Hale Maral Kır, Bülent Şık
Scandinavian Journal of Clinical and Laboratory Investigation
Bacterial Identification and Susceptibility Testing
article

Comparative evaluation of diagnostic and analytical performance of DIRUI FUS-200 and MUS-3600 fully automated urine analyzers

Fatih Hunc, Berrin Öztaş, Hale Maral Kır, Bülent Şık
article en

Abstract

The aim was to assess and compare the analytical performance of the FUS-200 and its new upgrade, the MUS-3600 urine autoanalyzers, and to compare performance with manual microscopy. First morning, void urine samples were randomly collected, and suitable samples were analyzed on both autoanalyzers. In addition, concurrent manual microscopic examinations were performed for all suitable urine samples. Carry-over, linearity and imprecision analysis were performed to assess analytical and diagnostic performance of both urine autoanalyzers according to Clinical and Laboratory Standards Institute (CLSI) EP15-A2, and the study was conducted in accordance with CLSI GP16-A3 Urinalysis; Approved Guideline. A total of 518 samples were collected, and of these, 494 (95.4%) were suitable for analysis and were included in the study. Sensitivity and positive likelihood ratio (LR+) values for WBC and squamous epithelium (SqEC) counts for both autoanalyzers were >90% and >10%, respectively. Specificity and LR+ values obtained from MUS-3600 were better compared to the FUS-200 (respectively 92.8 vs. 84.7; 5.5 vs. 10.1). One group agreement between the MUS-3600 and manual microscopy was >90%. Both analyzers displayed comparable performance for RBC and WBC counts, which were moderately correlated with each other. These results show the diagnostic performance of the MUS-3600 and FUS-200 was satisfactory for urine sediment analysis and was compatible with manual microscopy findings. However, random urine samples are sub-optimal for evaluating diagnostic performance, particularly for bacterial, cast, crystal and yeast analysis. Thus, we recommend using more appropriate urine samples for this comparison, such as from patients with suspected urinary tract infections.

Scandinavian Journal of Clinical and Laboratory InvestigationVol. 85(7)
Kocaeli Üniversitesi (TR)
Openalex Percentile: Top 63%
Bacterial Identification and Susceptibility Testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparative evaluation of diagnostic and analytical performance of DIRUI FUS-200 and MUS-3600 fully automated urine analyzers — Fatih Hunc, Berrin Öztaş, et al. · Scandinavian Journal of Clinical and Laboratory Investigation (2025) | TGRS Research Map | TGRS