Evaluation of the accuracy of intraoral scanners for nasal digitization: an in vitro three-dimensional comparative study
Taking a facial impression is essential in maxillofacial prosthetics and in dentistry when designing a dental prosthesis. This study explores the use of intraoral scanners as an alternative method for recording the surface of the face, specifically the nose. A soft head model was scanned using three different intraoral scanners, and the accuracy with which they captured the nose was compared with that obtained by conventional impressions. Two of the three scanners successfully captured the nose, demonstrating trueness and precision superior to that of conventional impressions. Intraoral scanners are a viable option for capturing the surface of the nose.
Authors
- Sandra Stadler-Heer (ORCID: https://orcid.org/0000-0003-3881-0984)
- Ralf‐Joachim Kohal (ORCID: https://orcid.org/0000-0001-7095-4190)
- Mahmoud Elbashti (ORCID: https://orcid.org/0000-0002-0215-5335)
- Yuka I. Sumita (ORCID: https://orcid.org/0000-0003-3982-8369)
- Mariko Hattori (ORCID: https://orcid.org/0000-0003-2802-8016)
- Noriyuki Wakabayashi (ORCID: https://orcid.org/0000-0002-0517-6756)
- Kirstin Vach (ORCID: https://orcid.org/0000-0001-9278-2203)
Institutions
- Tokyo Medical and Dental University (JP)
- University of Freiburg (DE)
- University Medical Center Freiburg (DE)
- Al Zawiya University (LY)
- Nippon Dental University (JP)
- Oras (Germany) (DE)
Publication Details
- Journal
- Frontiers in Oral Health
- Published
- 2025-11-07
- DOI
- https://doi.org/10.3389/froh.2025.1678778
- Primary Topic
- Reconstructive Facial Surgery Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Tokyo Metropolitan Government
- Albert-Ludwigs-Universität Freiburg
- Japan Society for the Promotion of Science