Anatomically Guided Motion Correction for Placental IVIM Parameter Estimation With Accelerated Sampling Method

Intravoxel incoherent motion (IVIM) is a diffusion-weighted magnetic resonance imaging (MRI) method that may be applied to the placenta to help diagnose abnormal pregnancies. IVIM requires prolonged scan times, followed by a model-based estimation procedure. Maternal or fetal motion during the scan affects the accuracy of this estimation. In this work, we proposed to address this challenging motion correction and data fitting problem by using additional anatomical information that is routinely collected at the beginning of the examination. Super-resolution reconstruction (SRR) was applied to these anatomical data, to provide a patient-specific, 3D isotropic, anatomic reference. Our first contribution is a novel framework with a two-step motion correction that uses both IVIM and the SRR anatomic data, accounting for both intra- and inter-scan, non-rigid motion. Our second contribution is an automation and acceleration of the IVIM data fitting, using a state-of-the-art Bayesian-type algorithm, modified with a preconditioned Crank-Nicholson (pCN) sampling strategy. The accuracy of the IVIM parameter fitting was improved by the proposed motion correction strategy, as assessed by the mean absolute fitting error in the region of interest, which was 4.14 before and 3.02 after correction (arbitrary units of signal intensity). The novel sampling strategy accelerated parameter estimation by 39% in average, with the same accuracy as that of the conventional Bayesian approach. In conclusion, the proposed method may be applied to obtain fast and reliable IVIM parameter estimates in challenging scenarios such as prenatal MRI.

Authors

Institutions

Publication Details

Journal
IEEE Access
Published
2025-01-01
DOI
https://doi.org/10.1109/access.2025.3629631
Primary Topic
MRI in cancer diagnosis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Anatomically Guided Motion Correction for Placental IVIM Parameter Estimation With Accelerated Sampling Method

Bailiang Chen, Freddy Odille, Marine Beaumont, Charline Bertholdt et al.
IEEE Access
MRI in cancer diagnosis
article

Anatomically Guided Motion Correction for Placental IVIM Parameter Estimation With Accelerated Sampling Method

Bailiang Chen, Freddy Odille, Marine Beaumont, Charline Bertholdt, Mbaimou Auxence Ngremmadji, Olivier Morel
article en

Abstract

Intravoxel incoherent motion (IVIM) is a diffusion-weighted magnetic resonance imaging (MRI) method that may be applied to the placenta to help diagnose abnormal pregnancies. IVIM requires prolonged scan times, followed by a model-based estimation procedure. Maternal or fetal motion during the scan affects the accuracy of this estimation. In this work, we proposed to address this challenging motion correction and data fitting problem by using additional anatomical information that is routinely collected at the beginning of the examination. Super-resolution reconstruction (SRR) was applied to these anatomical data, to provide a patient-specific, 3D isotropic, anatomic reference. Our first contribution is a novel framework with a two-step motion correction that uses both IVIM and the SRR anatomic data, accounting for both intra- and inter-scan, non-rigid motion. Our second contribution is an automation and acceleration of the IVIM data fitting, using a state-of-the-art Bayesian-type algorithm, modified with a preconditioned Crank-Nicholson (pCN) sampling strategy. The accuracy of the IVIM parameter fitting was improved by the proposed motion correction strategy, as assessed by the mean absolute fitting error in the region of interest, which was 4.14 before and 3.02 after correction (arbitrary units of signal intensity). The novel sampling strategy accelerated parameter estimation by 39% in average, with the same accuracy as that of the conventional Bayesian approach. In conclusion, the proposed method may be applied to obtain fast and reliable IVIM parameter estimates in challenging scenarios such as prenatal MRI.

IEEE AccessVol. 13
Inserm (FR), Imagerie Adaptative Diagnostique et Interventionnelle (FR), Université de Lorraine (FR)
Openalex Percentile: Top 52%
MRI in cancer diagnosis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.