Simultaneous OCT A-scan depth ranging and C-scan volume imaging for intraoperative robotic retinal surgery guidance

Current subretinal injection systems combine surgical microscopy, optical coherence tomography (OCT), and robotic assistance to enable precise, depth-resolved, and tremor-free drug delivery to targeted retinal sites. Nevertheless, subretinal and intraretinal vascular injections remain constrained by two major challenges: (1) the inherent difficulty in simultaneously achieving wide-field, real-time imaging of fundus lesions at high spatial resolution, and (2) the requirement for precise intraocular localization of the needle tip quickly. Current fundus injection imaging methods provide limited accuracy in needle tip localization and require prolonged procedure time. To address this issue, we present a novel intraoperative OCT (iOCT) imaging system that integrates a single light source and dual-channel OCT detection to simultaneously acquire A-scan and C-scan data without incurring substantial additional cost. Using the A-scan component of the synchronization signal, high-speed and real-time distance detection of the retinal surface is performed in M-scan mode. Meanwhile, the C-scan component of the synchronization signal provides high spatial resolution imaging of retinal depth information. Combined with the ultra-wide field of view afforded by microscopic imaging, this approach facilitates rapid and extensive guidance for lesion localization in the fundus, while also allowing real-time intraoperative assessment of drug injection procedures. This system is integrated into a high-precision robot-assisted platform, enabling subretinal injections and retinal vascular injections within the ocular fundus. To the best of our knowledge, this is the first time that the simultaneous acquisition of A-scan and C-scan data has been proposed for subretinal injection and retinal vascular injection procedures. The A-scan sensitivity roll-off was characterized and exhibited a decrease of approximately 3 dB over a 0.88 mm depth range in water. The B-scan imaging with 1024 × 1024 pixels achieve a 97.7 fps with maximal lateral field of view (FOV) 16 mm. The system achieved an axial measurement accuracy of 15 µm for needle tip localization. The success rate of single attempt subretinal (6 out of 6 attempts) and optic disc vessel injections was 100% in both live (3 out of 3 attempts) and <i>ex vivo</i> pig eyes (100 out of 100 attempts/30 out of 30 attempts), while venous injections in live pig eyes required positional adjustments during injection to achieve a 100% success rate (3 out of 3 attempts). Integrated with a microscope, the A-scan and C-scan OCT system, assisted by a robotic arm, can safely and reliably address the challenges associated with subretinal injection and retinal vascular injection.

Authors

Institutions

Publication Details

Journal
Biomedical Optics Express
Published
2025-11-06
DOI
https://doi.org/10.1364/boe.581579
Primary Topic
Retinal and Macular Surgery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simultaneous OCT A-scan depth ranging and C-scan volume imaging for intraoperative robotic retinal surgery guidance

Yujie Bai, Yuanyuan Deng, Pan Fu, Yougang Zhai et al.
Biomedical Optics Express
Retinal and Macular Surgery
article

Simultaneous OCT A-scan depth ranging and C-scan volume imaging for intraoperative robotic retinal surgery guidance

Yujie Bai, Yuanyuan Deng, Pan Fu, Yougang Zhai, Gui-Bin Bian, Boyu Yang, Yingxiong Xie, Yong Huang, Qun Hao
article en

Abstract

Current subretinal injection systems combine surgical microscopy, optical coherence tomography (OCT), and robotic assistance to enable precise, depth-resolved, and tremor-free drug delivery to targeted retinal sites. Nevertheless, subretinal and intraretinal vascular injections remain constrained by two major challenges: (1) the inherent difficulty in simultaneously achieving wide-field, real-time imaging of fundus lesions at high spatial resolution, and (2) the requirement for precise intraocular localization of the needle tip quickly. Current fundus injection imaging methods provide limited accuracy in needle tip localization and require prolonged procedure time. To address this issue, we present a novel intraoperative OCT (iOCT) imaging system that integrates a single light source and dual-channel OCT detection to simultaneously acquire A-scan and C-scan data without incurring substantial additional cost. Using the A-scan component of the synchronization signal, high-speed and real-time distance detection of the retinal surface is performed in M-scan mode. Meanwhile, the C-scan component of the synchronization signal provides high spatial resolution imaging of retinal depth information. Combined with the ultra-wide field of view afforded by microscopic imaging, this approach facilitates rapid and extensive guidance for lesion localization in the fundus, while also allowing real-time intraoperative assessment of drug injection procedures. This system is integrated into a high-precision robot-assisted platform, enabling subretinal injections and retinal vascular injections within the ocular fundus. To the best of our knowledge, this is the first time that the simultaneous acquisition of A-scan and C-scan data has been proposed for subretinal injection and retinal vascular injection procedures. The A-scan sensitivity roll-off was characterized and exhibited a decrease of approximately 3 dB over a 0.88 mm depth range in water. The B-scan imaging with 1024 × 1024 pixels achieve a 97.7 fps with maximal lateral field of view (FOV) 16 mm. The system achieved an axial measurement accuracy of 15 µm for needle tip localization. The success rate of single attempt subretinal (6 out of 6 attempts) and optic disc vessel injections was 100% in both live (3 out of 3 attempts) and <i>ex vivo</i> pig eyes (100 out of 100 attempts/30 out of 30 attempts), while venous injections in live pig eyes required positional adjustments during injection to achieve a 100% success rate (3 out of 3 attempts). Integrated with a microscope, the A-scan and C-scan OCT system, assisted by a robotic arm, can safely and reliably address the challenges associated with subretinal injection and retinal vascular injection.

Biomedical Optics ExpressVol. 16(12)
Beijing Institute of Technology (CN), Beijing Automation Control Equipment Institute (CN), University of Doha for Science and Technology (QA), Science and Technology on Surface Physics and Chemistry Laboratory (CN), Huanggang Central Hospital (CN), Taiyuan University of Technology (CN)
Openalex Percentile: Top 52%
Retinal and Macular Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.