Fast Neutron Multiplicity Counting for High Mass Plutonium Assay

The fast neutron multiplicity counting (FNMC) apparatus was consisted of 36 pieces of liquid scintillator detectors in three rings cylindrical layout, multi-modules of 500M14bit digitizers based on PXIe platform, and mechanical structure. Its center cavity was 35 cm in radius and 50 cm in height for assaying the plutonium item in standard container. The digitizers implemented detector’s signal acquisition, real-time digital data processing and data throughput. Each detector’s light output was calibration with Cs-137 source for nearly equal response. The apparatus detection efficiency for fast neutron calibrated with Cf-252 source was about 3.2%. The fast neutron multiplicity counting analysis was verified with Cf-252 neutron source, then experiments were performed with a serial of different high mass plutonium metals. Two-parameter calibration procedure was used for mass resolving. The mass assay results were consistent with the sample nominal values within maximum bias of 6%. This research achievement demonstrated a promising application for nuclear material assay and verification.

Authors

Institutions

Publication Details

Journal
EPJ Web of Conferences
Published
2025-01-01
DOI
https://doi.org/10.1051/epjconf/202533806012
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fast Neutron Multiplicity Counting for High Mass Plutonium Assay

X.T. Liu, Jiansheng Li, Yu Jin
EPJ Web of Conferences
Radiation Detection and Scintillator Technologies
article

Fast Neutron Multiplicity Counting for High Mass Plutonium Assay

X.T. Liu, Jiansheng Li, Yu Jin
article en

Abstract

The fast neutron multiplicity counting (FNMC) apparatus was consisted of 36 pieces of liquid scintillator detectors in three rings cylindrical layout, multi-modules of 500M14bit digitizers based on PXIe platform, and mechanical structure. Its center cavity was 35 cm in radius and 50 cm in height for assaying the plutonium item in standard container. The digitizers implemented detector’s signal acquisition, real-time digital data processing and data throughput. Each detector’s light output was calibration with Cs-137 source for nearly equal response. The apparatus detection efficiency for fast neutron calibrated with Cf-252 source was about 3.2%. The fast neutron multiplicity counting analysis was verified with Cf-252 neutron source, then experiments were performed with a serial of different high mass plutonium metals. Two-parameter calibration procedure was used for mass resolving. The mass assay results were consistent with the sample nominal values within maximum bias of 6%. This research achievement demonstrated a promising application for nuclear material assay and verification.

EPJ Web of ConferencesVol. 338
Institute of Nuclear Physics (UZ)
Openalex Percentile: Top 57%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.