New particle therapy modalities, including ultra-high dose rate (UHDR) proton therapy, present various challenges for particle detectors used in patient-specific quality assurance (PSQA). Amorphous Silicon (a-Si) Flat Panel (FP) detectors are a powerful tool used to create a 2D image of an administered dose map. They are known for high radiation hardness and image resolution. To investigate and characterize the behavior of a-Si pixels under particle therapy irradiation, specifically proton irradiation, a 2.4 x 2.4 mm pixel array of a-Si pixels, photodiodes coupled with thin-film-transistors (TFTs), was acquired and tested. Before irradiation, electronic probing was performed to generate the I-V curves and to qualify the stability of the pixel current for different bias and gate voltages to understand the proper voltage regime for optimal pixel functionality. In addition, proton irradiation was performed using pencil beam scanning (PBS) in a medical facility for charge linearity verification through direct detection. The test was done in conjunction with an IBA Razor TM Detector to compare each detector’s performance. Finally, irradiation with a high-power green laser was performed to examine the performance of the individual components of the pixels under high-intensity irradiation.

Authors

Institutions

Publication Details

Journal
EPJ Web of Conferences
Published
2025-01-01
DOI
https://doi.org/10.1051/epjconf/202533809008
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility study of a 2D Amorphous Silicon-based Detector for Proton Therapy

Séverine Rossomme, E. Cortina, Khalil El Achi, Alexis Warnier et al.
EPJ Web of Conferences
Radiation Therapy and Dosimetry
article

Feasibility study of a 2D Amorphous Silicon-based Detector for Proton Therapy

Séverine Rossomme, E. Cortina, Khalil El Achi, Alexis Warnier, Victor de Beco
article en

Abstract

New particle therapy modalities, including ultra-high dose rate (UHDR) proton therapy, present various challenges for particle detectors used in patient-specific quality assurance (PSQA). Amorphous Silicon (a-Si) Flat Panel (FP) detectors are a powerful tool used to create a 2D image of an administered dose map. They are known for high radiation hardness and image resolution. To investigate and characterize the behavior of a-Si pixels under particle therapy irradiation, specifically proton irradiation, a 2.4 x 2.4 mm pixel array of a-Si pixels, photodiodes coupled with thin-film-transistors (TFTs), was acquired and tested. Before irradiation, electronic probing was performed to generate the I-V curves and to qualify the stability of the pixel current for different bias and gate voltages to understand the proper voltage regime for optimal pixel functionality. In addition, proton irradiation was performed using pencil beam scanning (PBS) in a medical facility for charge linearity verification through direct detection. The test was done in conjunction with an IBA Razor TM Detector to compare each detector’s performance. Finally, irradiation with a high-power green laser was performed to examine the performance of the individual components of the pixels under high-intensity irradiation.

EPJ Web of ConferencesVol. 338
MGS Research (United States) (US), UCLouvain (BE)
Openalex Percentile: Top 53%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.