Development of a transportable neutron spectrometer for operational dosimetry within the DONEUT project

Accurate neutron dose rate estimation demands precise measurement of both kinetic energy and neutron flux. This study introduces a multi-instrumented transportable neutron spectrometer/dose rate meter, developed as part of the DONEUT project. The prototype, optimized for 0–20 MeV neutron dose rates, is based on a polyethylene moderator with thermal neutron detectors at different depths to reconstruct energy spectra via unfolding algorithms. The initial DONEUT V1 design consists of a 25 cm diameter polyethylene cylinder with 32 embedded ZnS:LiF scintillators, coupled to a SiPM matrix via light guides and wavelength-shifting fibers. GEANT4 Monte-Carlo simulations guided detector configuration and design optimization. Unfolding methods (MAXED and GRAVEL) were applied to various moderator volumes and neutron energy spectra, including Am/Be and 252 Cf sources. This led to a “dome” prototype, DONEUT V1+, which improves vertical angles response uniformity and reduces weight. A sixth detection depth was added to enhance deconvolution robustness, particularly in the epithermal range, while the number of detectors was reduced from 32 to 24. A custom Python implementation of the GRAVEL algorithm was developed, replacing traditional codes, increasing flexibility and processing speed for energy spectra reconstruction. Experimental tests with moderated and unmoderated Am/Be sources were performed, benchmarked against NNS measurements, demonstrated dose equivalent rate (H*(10)) accuracy within 15%, even at low dose rates (~1 µSv/h) over a 10-minute measurement. The prototype reliably assesses dose rates from 1 to 100 µSv/h. A new version, DONEUTV2, insensitive to gamma background, is under development to extend the range to several mSv/h

Authors

Institutions

Publication Details

Journal
EPJ Web of Conferences
Published
2025-01-01
DOI
https://doi.org/10.1051/epjconf/202533810011
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a transportable neutron spectrometer for operational dosimetry within the DONEUT project

G Bouteiller, J. L. Trolet, Claude-Alexandre Simonetti, M. Labalme et al.
EPJ Web of Conferences
Radiation Therapy and Dosimetry
article

Development of a transportable neutron spectrometer for operational dosimetry within the DONEUT project

G Bouteiller, J. L. Trolet, Claude-Alexandre Simonetti, M. Labalme, Alexandre Roussel, S. Bozec, Y. Brelet, A. Hemmer, E. Moncanis
article en

Abstract

Accurate neutron dose rate estimation demands precise measurement of both kinetic energy and neutron flux. This study introduces a multi-instrumented transportable neutron spectrometer/dose rate meter, developed as part of the DONEUT project. The prototype, optimized for 0–20 MeV neutron dose rates, is based on a polyethylene moderator with thermal neutron detectors at different depths to reconstruct energy spectra via unfolding algorithms. The initial DONEUT V1 design consists of a 25 cm diameter polyethylene cylinder with 32 embedded ZnS:LiF scintillators, coupled to a SiPM matrix via light guides and wavelength-shifting fibers. GEANT4 Monte-Carlo simulations guided detector configuration and design optimization. Unfolding methods (MAXED and GRAVEL) were applied to various moderator volumes and neutron energy spectra, including Am/Be and 252 Cf sources. This led to a “dome” prototype, DONEUT V1+, which improves vertical angles response uniformity and reduces weight. A sixth detection depth was added to enhance deconvolution robustness, particularly in the epithermal range, while the number of detectors was reduced from 32 to 24. A custom Python implementation of the GRAVEL algorithm was developed, replacing traditional codes, increasing flexibility and processing speed for energy spectra reconstruction. Experimental tests with moderated and unmoderated Am/Be sources were performed, benchmarked against NNS measurements, demonstrated dose equivalent rate (H*(10)) accuracy within 15%, even at low dose rates (~1 µSv/h) over a 10-minute measurement. The prototype reliably assesses dose rates from 1 to 100 µSv/h. A new version, DONEUTV2, insensitive to gamma background, is under development to extend the range to several mSv/h

EPJ Web of ConferencesVol. 338
Centre National de la Recherche Scientifique (FR), École Nationale Supérieure d'Ingénieurs de Caen (FR), Normandie Université (FR), Laboratoire de Physique Corpusculaire de Caen (FR), École Nationale d'Administration et de Magistrature (NE), Université de Caen Normandie (FR)
Région Normandie
Openalex Percentile: Top 53%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.