Subcutaneous and continuous blood pressure monitoring in an ambulatory sheep by piezoelectric micromachined ultrasonic transducers

This paper presents subcutaneous and continuous blood pressure (BP) monitoring using aluminum nitride (AlN) piezoelectric micromachined ultrasonic transducers (PMUTs) in an ambulatory sheep. A 37 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>×</mml:mo></mml:math> 45 PMUTs array with a footprint of 5 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>×</mml:mo></mml:math> 5 mm<sup>2</sup> has been designed and fabricated as a prototype device. The deep reactive ion etching (DRIE) process to open the backside holes on the silicon substrate has been optimized to create active device diaphragms with a radius of 29 μm. The resulting PMUT unit has a measured resonant frequency of 6.5 MHz in water, an output acoustic pressure of 28 kPa at a distance of 10 mm, and a 6-dB bandwidth of about 33%. The BP monitoring scheme is validated through both in vitro and in vivo experiments to illustrate the correlation between the diameter of the blood vessel and pressure. Simulations indicate that possible issues in misalignment between the device and the blood vessel can result in a 60% reduction in signal strength with only 1 mm in misalignment. This highlights the advantage of subcutaneous implantation in maintaining a stable interface and consistent alignment for reliable long-term BP monitoring, in contrast to similar approaches via wearable system setups. The in vivo testing result shows BP wave fine features such as dicrotic notches and the averaged systolic/diastolic pressure errors are -1.2 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>±</mml:mo></mml:math> 2.1 and -2.9 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>±</mml:mo></mml:math> 1.4 mmHg, respectively, which meets the clinical standard as calibrated by a gold-standard arterial line pressure sensor. As such, this system highlights the potential applications in silent, continuous, and highly accurate BP monitoring for hypertension patients using this implantable MEMS-based technology.

Authors

Institutions

Publication Details

Journal
Microsystems & Nanoengineering
Published
2025-11-06
DOI
https://doi.org/10.1038/s41378-025-01019-w
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Subcutaneous and continuous blood pressure monitoring in an ambulatory sheep by piezoelectric micromachined ultrasonic transducers

Zhichun Shao, Hong Ding, Yande Peng, Wei Yue et al.
Microsystems & Nanoengineering
Ultrasound Imaging and Elastography
article

Subcutaneous and continuous blood pressure monitoring in an ambulatory sheep by piezoelectric micromachined ultrasonic transducers

Zhichun Shao, Hong Ding, Yande Peng, Wei Yue, Jin Xie, Fan Xia, Sedat Pala, Liwei Lin
article en

Abstract

This paper presents subcutaneous and continuous blood pressure (BP) monitoring using aluminum nitride (AlN) piezoelectric micromachined ultrasonic transducers (PMUTs) in an ambulatory sheep. A 37 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>×</mml:mo></mml:math> 45 PMUTs array with a footprint of 5 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>×</mml:mo></mml:math> 5 mm<sup>2</sup> has been designed and fabricated as a prototype device. The deep reactive ion etching (DRIE) process to open the backside holes on the silicon substrate has been optimized to create active device diaphragms with a radius of 29 μm. The resulting PMUT unit has a measured resonant frequency of 6.5 MHz in water, an output acoustic pressure of 28 kPa at a distance of 10 mm, and a 6-dB bandwidth of about 33%. The BP monitoring scheme is validated through both in vitro and in vivo experiments to illustrate the correlation between the diameter of the blood vessel and pressure. Simulations indicate that possible issues in misalignment between the device and the blood vessel can result in a 60% reduction in signal strength with only 1 mm in misalignment. This highlights the advantage of subcutaneous implantation in maintaining a stable interface and consistent alignment for reliable long-term BP monitoring, in contrast to similar approaches via wearable system setups. The in vivo testing result shows BP wave fine features such as dicrotic notches and the averaged systolic/diastolic pressure errors are -1.2 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>±</mml:mo></mml:math> 2.1 and -2.9 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mo>±</mml:mo></mml:math> 1.4 mmHg, respectively, which meets the clinical standard as calibrated by a gold-standard arterial line pressure sensor. As such, this system highlights the potential applications in silent, continuous, and highly accurate BP monitoring for hypertension patients using this implantable MEMS-based technology.

Microsystems & NanoengineeringVol. 11(1)
University of California San Diego (US), Zhejiang University (CN), University of California, Berkeley (US)
University of Minnesota, British Society for Antimicrobial Chemotherapy
Openalex Percentile: Top 52%
Ultrasound Imaging and Elastography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.